A randomised RSA study of Peri-Apatitetrade mark HA coating of a total knee prosthesis.
(2008) In Knee 15. p.211-216- Abstract
- BACKGROUND: In total knee replacement, sound early fixation of the prosthesis is crucial for achieving a good long-term result and for minimising the risk of loosening. Various types of prosthetic material, different surface textures and shapes and the incorporation of screws or pegs have been used to achieve good fixation, particularly in the uncemented knee. Hydroxyapatite (HA) coating of prosthetic joint components is another technique used to enhance early stability and so to improve the longevity of the prosthesis-bone fixation. HA ceramic coatings are mostly plasma sprayed onto the fixation surface of the implant. Plasma spraying is largely a 'line of sight' technique and as such there are difficulties involved when covering... (More)
- BACKGROUND: In total knee replacement, sound early fixation of the prosthesis is crucial for achieving a good long-term result and for minimising the risk of loosening. Various types of prosthetic material, different surface textures and shapes and the incorporation of screws or pegs have been used to achieve good fixation, particularly in the uncemented knee. Hydroxyapatite (HA) coating of prosthetic joint components is another technique used to enhance early stability and so to improve the longevity of the prosthesis-bone fixation. HA ceramic coatings are mostly plasma sprayed onto the fixation surface of the implant. Plasma spraying is largely a 'line of sight' technique and as such there are difficulties involved when covering three-dimensional planes, such as porous beaded fixation surfaces typically found on several knee prostheses. The objective of this study was to assess the clinical performance of the solution-deposited HA coating, Peri-Apatitetrade mark , with regard to its ability to stimulate an endurable and stable implant fixation. PATIENTS AND METHODS: We randomised 60 patients into two groups; one group received the porous coated prosthesis with solution-deposited HA, and the other group received a prosthesis without HA. Radiostereometric examination was used as the primary tool for measuring migration in the prosthetic components. RESULTS: There was a lower incidence of early subsidence in the Peri-Apatitetrade mark group. At 24 months there were no differences in clinical scorings or maximal total point motion. CONCLUSION: Addition of solution-deposited HA coating appears to provide better early stable fixation in a porous coated knee prosthesis. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1052703
- author
- Hansson, Ulrik LU ; Ryd, Leif LU and Toksvig-Larsen, Sören LU
- organization
- publishing date
- 2008
- type
- Contribution to journal
- publication status
- published
- subject
- in
- Knee
- volume
- 15
- pages
- 211 - 216
- publisher
- Elsevier
- external identifiers
-
- pmid:18329882
- wos:000256749100012
- scopus:42749089229
- pmid:18329882
- ISSN
- 1873-5800
- DOI
- 10.1016/j.knee.2008.01.013
- language
- English
- LU publication?
- yes
- id
- 1aaa7b7b-a595-486b-be71-7dbc1564d2b8 (old id 1052703)
- alternative location
- http://www.ncbi.nlm.nih.gov/pubmed/18329882?dopt=Abstract
- date added to LUP
- 2016-04-04 07:11:00
- date last changed
- 2022-01-29 01:49:40
@article{1aaa7b7b-a595-486b-be71-7dbc1564d2b8, abstract = {{BACKGROUND: In total knee replacement, sound early fixation of the prosthesis is crucial for achieving a good long-term result and for minimising the risk of loosening. Various types of prosthetic material, different surface textures and shapes and the incorporation of screws or pegs have been used to achieve good fixation, particularly in the uncemented knee. Hydroxyapatite (HA) coating of prosthetic joint components is another technique used to enhance early stability and so to improve the longevity of the prosthesis-bone fixation. HA ceramic coatings are mostly plasma sprayed onto the fixation surface of the implant. Plasma spraying is largely a 'line of sight' technique and as such there are difficulties involved when covering three-dimensional planes, such as porous beaded fixation surfaces typically found on several knee prostheses. The objective of this study was to assess the clinical performance of the solution-deposited HA coating, Peri-Apatitetrade mark , with regard to its ability to stimulate an endurable and stable implant fixation. PATIENTS AND METHODS: We randomised 60 patients into two groups; one group received the porous coated prosthesis with solution-deposited HA, and the other group received a prosthesis without HA. Radiostereometric examination was used as the primary tool for measuring migration in the prosthetic components. RESULTS: There was a lower incidence of early subsidence in the Peri-Apatitetrade mark group. At 24 months there were no differences in clinical scorings or maximal total point motion. CONCLUSION: Addition of solution-deposited HA coating appears to provide better early stable fixation in a porous coated knee prosthesis.}}, author = {{Hansson, Ulrik and Ryd, Leif and Toksvig-Larsen, Sören}}, issn = {{1873-5800}}, language = {{eng}}, pages = {{211--216}}, publisher = {{Elsevier}}, series = {{Knee}}, title = {{A randomised RSA study of Peri-Apatitetrade mark HA coating of a total knee prosthesis.}}, url = {{http://dx.doi.org/10.1016/j.knee.2008.01.013}}, doi = {{10.1016/j.knee.2008.01.013}}, volume = {{15}}, year = {{2008}}, }