Modeling transmission and scatter for photon beam attenuators

Ahnesjö, Anders ; Weber, Lars LU and Nilsson, Per LU orcid (1995) In Medical Physics 22(11). p.1711-1720
Abstract
The development of treatment planning methods in radiation therapy requires dose calculation methods that are both accurate and general enough to provide a dose per unit monitor setting for a broad variety of fields and beam modifiers. The purpose of this work was to develop models for calculation of scatter and transmission for photon beam attenuators such as compensating filters, wedges, and block trays. The attenuation of the beam is calculated using a spectrum of the beam, and a correction factor based on attenuation measurements. Small angle coherent scatter and electron binding effects on scattering cross sections are considered by use of a correction factor. Quality changes in beam penetrability and energy fluence to dose conversion... (More)
The development of treatment planning methods in radiation therapy requires dose calculation methods that are both accurate and general enough to provide a dose per unit monitor setting for a broad variety of fields and beam modifiers. The purpose of this work was to develop models for calculation of scatter and transmission for photon beam attenuators such as compensating filters, wedges, and block trays. The attenuation of the beam is calculated using a spectrum of the beam, and a correction factor based on attenuation measurements. Small angle coherent scatter and electron binding effects on scattering cross sections are considered by use of a correction factor. Quality changes in beam penetrability and energy fluence to dose conversion are modeled by use of the calculated primary beam spectrum after passage through the attenuator. The beam spectra are derived by the depth dose effective method, i.e., by minimizing the difference between measured and calculated depth dose distributions, where the calculated distributions are derived by superposing data from a database for monoenergetic photons. The attenuator scatter is integrated over the area viewed from the calculation point of view using first scatter theory. Calculations are simplified by replacing the energy and angular-dependent cross-section formulas with the forward scatter constant r2(0) and a set of parametrized correction functions. The set of corrections include functions for the Compton energy loss, scatter attenuation, and secondary bremsstrahlung production. The effect of charged particle contamination is bypassed by avoiding use of dmax for absolute dose calibrations. The results of the model are compared with scatter measurements in air for copper and lead filters and with dose to a water phantom for lead filters for 4 and 18 MV. For attenuated beams, downstream of the buildup region, the calculated results agree with measurements on the 1.5% level. The accuracy was slightly less in situations where the scatter component is very large, as for very large fields with very short filter to detector distances. The implementation of the model into treatment planning systems is discussed. (Less)
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Medical Physics
volume
22
issue
11
pages
1711 - 1720
publisher
American Association of Physicists in Medicine
external identifiers
  • pmid:8587523
  • scopus:0028783999
  • pmid:8587523
ISSN
0094-2405
DOI
10.1118/1.597534
language
English
LU publication?
yes
id
6dcb383d-7cc1-4612-95ad-f83de5612e3f (old id 1109004)
alternative location
http://dx.doi.org/10.1118/1.597534
date added to LUP
2016-04-01 16:06:11
date last changed
2021-01-03 05:35:14
@article{6dcb383d-7cc1-4612-95ad-f83de5612e3f,
  abstract     = {{The development of treatment planning methods in radiation therapy requires dose calculation methods that are both accurate and general enough to provide a dose per unit monitor setting for a broad variety of fields and beam modifiers. The purpose of this work was to develop models for calculation of scatter and transmission for photon beam attenuators such as compensating filters, wedges, and block trays. The attenuation of the beam is calculated using a spectrum of the beam, and a correction factor based on attenuation measurements. Small angle coherent scatter and electron binding effects on scattering cross sections are considered by use of a correction factor. Quality changes in beam penetrability and energy fluence to dose conversion are modeled by use of the calculated primary beam spectrum after passage through the attenuator. The beam spectra are derived by the depth dose effective method, i.e., by minimizing the difference between measured and calculated depth dose distributions, where the calculated distributions are derived by superposing data from a database for monoenergetic photons. The attenuator scatter is integrated over the area viewed from the calculation point of view using first scatter theory. Calculations are simplified by replacing the energy and angular-dependent cross-section formulas with the forward scatter constant r2(0) and a set of parametrized correction functions. The set of corrections include functions for the Compton energy loss, scatter attenuation, and secondary bremsstrahlung production. The effect of charged particle contamination is bypassed by avoiding use of dmax for absolute dose calibrations. The results of the model are compared with scatter measurements in air for copper and lead filters and with dose to a water phantom for lead filters for 4 and 18 MV. For attenuated beams, downstream of the buildup region, the calculated results agree with measurements on the 1.5% level. The accuracy was slightly less in situations where the scatter component is very large, as for very large fields with very short filter to detector distances. The implementation of the model into treatment planning systems is discussed.}},
  author       = {{Ahnesjö, Anders and Weber, Lars and Nilsson, Per}},
  issn         = {{0094-2405}},
  language     = {{eng}},
  number       = {{11}},
  pages        = {{1711--1720}},
  publisher    = {{American Association of Physicists in Medicine}},
  series       = {{Medical Physics}},
  title        = {{Modeling transmission and scatter for photon beam attenuators}},
  url          = {{http://dx.doi.org/10.1118/1.597534}},
  doi          = {{10.1118/1.597534}},
  volume       = {{22}},
  year         = {{1995}},
}