Validation of a 3D CT method for measurement of linear wear of acetabular cups.

Jedenmalm, Anneli LU ; Nilsson, Fritjof ; Noz, Marilyn E ; Green, Douglas D ; Gedde, Ulf W ; Clarke, Ian C ; Stark, Andreas ; Maguire, Gerald Q ; Zeleznik, Michael P and Olivecrona, Henrik (2011) In Acta Orthopaedica 82(1). p.35-41
Abstract
Background We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. Material and methods Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip simulator. Before and after wear, they were (1) imaged with a CT scanner using a phantom model device, (2) measured using a coordinate measurement machine (CMM), and (3) weighed. CMM was used as the reference method for measurement of femoral head penetration into the cup and for comparison with CT, and gravimetric measurements were used as a reference... (More)
Background We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. Material and methods Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip simulator. Before and after wear, they were (1) imaged with a CT scanner using a phantom model device, (2) measured using a coordinate measurement machine (CMM), and (3) weighed. CMM was used as the reference method for measurement of femoral head penetration into the cup and for comparison with CT, and gravimetric measurements were used as a reference for both CT and CMM. Femoral head penetration and wear vector angle were studied. The head diameters were also measured with both CMM and CT. The repeatability of the method proposed was evaluated with two repeated measurements using different positions of the phantom in the CT scanner. Results The accuracy of the 3D CT method for evaluation of linear wear was 0.51 mm and the repeatability was 0.39 mm. Repeatability for wear vector angle was 17?. Interpretation This study of metal-meshed hip-simulated acetabular cups shows that CT has the capacity for reliable measurement of linear wear of acetabular cups at a clinically relevant level of accuracy. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Acta Orthopaedica
volume
82
issue
1
pages
35 - 41
publisher
Taylor & Francis
external identifiers
  • wos:000287025400006
  • pmid:21281259
  • scopus:79751483254
  • pmid:21281259
ISSN
1745-3682
DOI
10.3109/17453674.2011.552777
language
English
LU publication?
yes
id
fa785d84-7bf9-4bff-b0d8-8e64bc5a980a (old id 1832494)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/21281259?dopt=Abstract
date added to LUP
2016-04-01 14:01:22
date last changed
2022-01-27 22:27:55
@article{fa785d84-7bf9-4bff-b0d8-8e64bc5a980a,
  abstract     = {{Background We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. Material and methods Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip simulator. Before and after wear, they were (1) imaged with a CT scanner using a phantom model device, (2) measured using a coordinate measurement machine (CMM), and (3) weighed. CMM was used as the reference method for measurement of femoral head penetration into the cup and for comparison with CT, and gravimetric measurements were used as a reference for both CT and CMM. Femoral head penetration and wear vector angle were studied. The head diameters were also measured with both CMM and CT. The repeatability of the method proposed was evaluated with two repeated measurements using different positions of the phantom in the CT scanner. Results The accuracy of the 3D CT method for evaluation of linear wear was 0.51 mm and the repeatability was 0.39 mm. Repeatability for wear vector angle was 17?. Interpretation This study of metal-meshed hip-simulated acetabular cups shows that CT has the capacity for reliable measurement of linear wear of acetabular cups at a clinically relevant level of accuracy.}},
  author       = {{Jedenmalm, Anneli and Nilsson, Fritjof and Noz, Marilyn E and Green, Douglas D and Gedde, Ulf W and Clarke, Ian C and Stark, Andreas and Maguire, Gerald Q and Zeleznik, Michael P and Olivecrona, Henrik}},
  issn         = {{1745-3682}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{35--41}},
  publisher    = {{Taylor & Francis}},
  series       = {{Acta Orthopaedica}},
  title        = {{Validation of a 3D CT method for measurement of linear wear of acetabular cups.}},
  url          = {{http://dx.doi.org/10.3109/17453674.2011.552777}},
  doi          = {{10.3109/17453674.2011.552777}},
  volume       = {{82}},
  year         = {{2011}},
}