Non-optimal focusing of ultrasound does not affect dual frequency ultrasound measurements

Malo, Markus ; Karjalainen, Janne ; Riekkinen, Ossi ; Isaksson, Hanna LU orcid ; Jurvelin, Jukka and Töyräs, Juha (2011) In IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control 58(6). p.1182-1188
Abstract
In pulse-echo (PE) ultrasound measurements, the use of focused transducers is desirable for quantitative assessment of bone characteristics because of the attenuation in the overlying soft tissues. However, the variable thickness and composition of the soft tissue overlying bone affect the focal depth of the ultrasound beam and induce errors into the measurements. To compensate for the attenuation-related effects caused by the interfering soft tissue (i.e., fat and lean tissue), a dual-frequency ultrasound (DFUS) technique was recently introduced. The aim of this study was to investigate the effect of non-optimal focal depth of the ultrasound beam on the determination of the integrated reflection coefficient (IRC) of bone when overlaid by... (More)
In pulse-echo (PE) ultrasound measurements, the use of focused transducers is desirable for quantitative assessment of bone characteristics because of the attenuation in the overlying soft tissues. However, the variable thickness and composition of the soft tissue overlying bone affect the focal depth of the ultrasound beam and induce errors into the measurements. To compensate for the attenuation-related effects caused by the interfering soft tissue (i.e., fat and lean tissue), a dual-frequency ultrasound (DFUS) technique was recently introduced. The aim of this study was to investigate the effect of non-optimal focal depth of the ultrasound beam on the determination of the integrated reflection coefficient (IRC) of bone when overlaid by an interfering layer composed of oil and water. The feasibility of the DFUS-based correction of the IRC was evaluated through numerical simulations and experimental measurements. Even when the interfering layer-bone interface was out of focus, the total thickness of the interfering layer could be accurately determined with the technique. However, based on the simulations, the errors in the determination of the composition of the interfering layer increased (0.004 to 113.8%) with the increase in distance between the interfering layer-bone interface and the focus of the ultrasound beam. Attenuation compensation, based on the true composition of the interfering layer, resulted in an average relative error of 22.3% in the IRC values calculated from the simulations. Interestingly, the attenuation compensation with the interfering layer composition estimated using the DFUS technique resulted in a smaller average relative error of 14.9% in the IRC values. The simulations suggest that DFUS can reduce the errors induced by soft tissue in bone PE ultrasound measurements. The experimental measurements indicate that the accuracy of the IRC measurements is rather similar when using DFUS correction or correction based on the true composition of - - the interfering layer. However, the results suggest that accurate determination of soft tissue composition may be difficult without optimal focusing of the ultrasound beam on the soft tissue-bone interface. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; and
publishing date
type
Contribution to journal
publication status
published
subject
in
IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
volume
58
issue
6
pages
1182 - 1188
publisher
IEEE - Institute of Electrical and Electronics Engineers Inc.
external identifiers
  • scopus:79959572209
ISSN
0885-3010
DOI
10.1109/TUFFC.2011.1928
language
English
LU publication?
no
id
0c871db2-26b8-4277-83b4-8de7f68779b6 (old id 2277253)
date added to LUP
2016-04-04 07:12:43
date last changed
2023-09-19 11:29:09
@article{0c871db2-26b8-4277-83b4-8de7f68779b6,
  abstract     = {{In pulse-echo (PE) ultrasound measurements, the use of focused transducers is desirable for quantitative assessment of bone characteristics because of the attenuation in the overlying soft tissues. However, the variable thickness and composition of the soft tissue overlying bone affect the focal depth of the ultrasound beam and induce errors into the measurements. To compensate for the attenuation-related effects caused by the interfering soft tissue (i.e., fat and lean tissue), a dual-frequency ultrasound (DFUS) technique was recently introduced. The aim of this study was to investigate the effect of non-optimal focal depth of the ultrasound beam on the determination of the integrated reflection coefficient (IRC) of bone when overlaid by an interfering layer composed of oil and water. The feasibility of the DFUS-based correction of the IRC was evaluated through numerical simulations and experimental measurements. Even when the interfering layer-bone interface was out of focus, the total thickness of the interfering layer could be accurately determined with the technique. However, based on the simulations, the errors in the determination of the composition of the interfering layer increased (0.004 to 113.8%) with the increase in distance between the interfering layer-bone interface and the focus of the ultrasound beam. Attenuation compensation, based on the true composition of the interfering layer, resulted in an average relative error of 22.3% in the IRC values calculated from the simulations. Interestingly, the attenuation compensation with the interfering layer composition estimated using the DFUS technique resulted in a smaller average relative error of 14.9% in the IRC values. The simulations suggest that DFUS can reduce the errors induced by soft tissue in bone PE ultrasound measurements. The experimental measurements indicate that the accuracy of the IRC measurements is rather similar when using DFUS correction or correction based on the true composition of - - the interfering layer. However, the results suggest that accurate determination of soft tissue composition may be difficult without optimal focusing of the ultrasound beam on the soft tissue-bone interface.}},
  author       = {{Malo, Markus and Karjalainen, Janne and Riekkinen, Ossi and Isaksson, Hanna and Jurvelin, Jukka and Töyräs, Juha}},
  issn         = {{0885-3010}},
  language     = {{eng}},
  number       = {{6}},
  pages        = {{1182--1188}},
  publisher    = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}},
  series       = {{IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control}},
  title        = {{Non-optimal focusing of ultrasound does not affect dual frequency ultrasound measurements}},
  url          = {{http://dx.doi.org/10.1109/TUFFC.2011.1928}},
  doi          = {{10.1109/TUFFC.2011.1928}},
  volume       = {{58}},
  year         = {{2011}},
}