Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

Response of Si p-i-n diode and Au/n-Si surface barrier detector to heavy ions

Zhang, Y W and Whitlow, Harry J LU (2002) In Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms 190(1-4). p.383-386
Abstract
Pulse height versus energy calibrations of a Si p-i-n diode and a Au-/n-Si surface barrier detector have been studied for heavy ions with atomic number (Z(1)) from 3 to 79 in a range from 0.1 to 0.8 MeV per nucleon as a function of bias voltage and detector tilting angle. The detector response is simultaneously measured using a time of flight-energy elastic recoil detection analysis set-up with recoils produced over a wide energy range from a thick target of each element. Prior to impinging on the Si detector, the individual recoil is tagged by its energy determined from the time of flight and tabulated isotopic mass. For both detectors, the pulse height-energy calibration for recoils with a given Z(1) is described well by a linear... (More)
Pulse height versus energy calibrations of a Si p-i-n diode and a Au-/n-Si surface barrier detector have been studied for heavy ions with atomic number (Z(1)) from 3 to 79 in a range from 0.1 to 0.8 MeV per nucleon as a function of bias voltage and detector tilting angle. The detector response is simultaneously measured using a time of flight-energy elastic recoil detection analysis set-up with recoils produced over a wide energy range from a thick target of each element. Prior to impinging on the Si detector, the individual recoil is tagged by its energy determined from the time of flight and tabulated isotopic mass. For both detectors, the pulse height-energy calibration for recoils with a given Z(1) is described well by a linear relationship with small systematic deviations. The linear-fit parameters show similar, but not identical dependence on both Z(1) and bias voltage (collecting field strength) for the surface barrier detectors and the p-i-n diode. These results suggest that the efficiency of electron-hole pair collection is markedly dependent on the different electric field configurations for the two detector structures. (Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
ToF-E ERDA, Si charged particle detector, surface barrier detector, energy calibration, electronic stopping
in
Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms
volume
190
issue
1-4
pages
383 - 386
publisher
Elsevier
external identifiers
  • wos:000176108800075
  • scopus:0036569791
ISSN
0168-583X
DOI
10.1016/S0168-583X(01)01240-X
language
English
LU publication?
yes
additional info
The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Nuclear Physics (Faculty of Technology) (011013007)
id
631f1602-aa52-43c8-9cce-20a638cf0b6c (old id 335447)
date added to LUP
2016-04-01 17:01:37
date last changed
2022-02-28 01:20:48
@article{631f1602-aa52-43c8-9cce-20a638cf0b6c,
  abstract     = {{Pulse height versus energy calibrations of a Si p-i-n diode and a Au-/n-Si surface barrier detector have been studied for heavy ions with atomic number (Z(1)) from 3 to 79 in a range from 0.1 to 0.8 MeV per nucleon as a function of bias voltage and detector tilting angle. The detector response is simultaneously measured using a time of flight-energy elastic recoil detection analysis set-up with recoils produced over a wide energy range from a thick target of each element. Prior to impinging on the Si detector, the individual recoil is tagged by its energy determined from the time of flight and tabulated isotopic mass. For both detectors, the pulse height-energy calibration for recoils with a given Z(1) is described well by a linear relationship with small systematic deviations. The linear-fit parameters show similar, but not identical dependence on both Z(1) and bias voltage (collecting field strength) for the surface barrier detectors and the p-i-n diode. These results suggest that the efficiency of electron-hole pair collection is markedly dependent on the different electric field configurations for the two detector structures.}},
  author       = {{Zhang, Y W and Whitlow, Harry J}},
  issn         = {{0168-583X}},
  keywords     = {{ToF-E ERDA; Si charged particle detector; surface barrier detector; energy calibration; electronic stopping}},
  language     = {{eng}},
  number       = {{1-4}},
  pages        = {{383--386}},
  publisher    = {{Elsevier}},
  series       = {{Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms}},
  title        = {{Response of Si p-i-n diode and Au/n-Si surface barrier detector to heavy ions}},
  url          = {{http://dx.doi.org/10.1016/S0168-583X(01)01240-X}},
  doi          = {{10.1016/S0168-583X(01)01240-X}},
  volume       = {{190}},
  year         = {{2002}},
}