The Duration of Negative Pressure Wound Therapy Can Be Reduced Using the HeartShield Device in Patients With Deep Sternal Wound Infection.

Ingemansson, Richard LU ; Malmsjö, Malin LU and Lindstedt Ingemansson, Sandra LU (2014) In Eplasty: Open Access Journal of Plastic and Reconstructive Surgery 14(Apr 3). p.16-16
Abstract
Background: Heart rupture resulting in lethal bleeding is a devastating complication associated with negative pressure wound therapy (NPWT) in patients with deep sternal wound infection (DSWI). We have previously reported that the use of a protective HeartShield device in combination with NPWT decreases the risk of damage to the heart. This article presents a retrospective analysis of NPWT duration with and without the HeartShield device. Subjects and patients: The study included 6 patients treated with the HeartShield device in combination with NPWT and 6 patients treated with conventional NPWT during the same time period. The duration of active treatment time was measured. Results: The median duration of NPWT was 8 days (range: 6-14... (More)
Background: Heart rupture resulting in lethal bleeding is a devastating complication associated with negative pressure wound therapy (NPWT) in patients with deep sternal wound infection (DSWI). We have previously reported that the use of a protective HeartShield device in combination with NPWT decreases the risk of damage to the heart. This article presents a retrospective analysis of NPWT duration with and without the HeartShield device. Subjects and patients: The study included 6 patients treated with the HeartShield device in combination with NPWT and 6 patients treated with conventional NPWT during the same time period. The duration of active treatment time was measured. Results: The median duration of NPWT was 8 days (range: 6-14 days) in the HeartShield device NPWT group and 14 days in the conventional group (range: 10-18 days). The difference was statistically significant (P < .05). Conclusions: It appears that the treatment of patients with DSWI with the HeartShield device reduces the duration of NPWT. (Less)
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Eplasty: Open Access Journal of Plastic and Reconstructive Surgery
volume
14
issue
Apr 3
pages
16 - 16
publisher
HMP Global Learning
external identifiers
  • pmid:24741387
ISSN
1937-5719
language
English
LU publication?
yes
id
ac781d97-e307-44a7-960d-96d0373e8f4b (old id 4429932)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/24741387?dopt=Abstract
date added to LUP
2016-04-04 09:14:37
date last changed
2018-11-21 20:51:46
@article{ac781d97-e307-44a7-960d-96d0373e8f4b,
  abstract     = {{Background: Heart rupture resulting in lethal bleeding is a devastating complication associated with negative pressure wound therapy (NPWT) in patients with deep sternal wound infection (DSWI). We have previously reported that the use of a protective HeartShield device in combination with NPWT decreases the risk of damage to the heart. This article presents a retrospective analysis of NPWT duration with and without the HeartShield device. Subjects and patients: The study included 6 patients treated with the HeartShield device in combination with NPWT and 6 patients treated with conventional NPWT during the same time period. The duration of active treatment time was measured. Results: The median duration of NPWT was 8 days (range: 6-14 days) in the HeartShield device NPWT group and 14 days in the conventional group (range: 10-18 days). The difference was statistically significant (P &lt; .05). Conclusions: It appears that the treatment of patients with DSWI with the HeartShield device reduces the duration of NPWT.}},
  author       = {{Ingemansson, Richard and Malmsjö, Malin and Lindstedt Ingemansson, Sandra}},
  issn         = {{1937-5719}},
  language     = {{eng}},
  number       = {{Apr 3}},
  pages        = {{16--16}},
  publisher    = {{HMP Global Learning}},
  series       = {{Eplasty: Open Access Journal of Plastic and Reconstructive Surgery}},
  title        = {{The Duration of Negative Pressure Wound Therapy Can Be Reduced Using the HeartShield Device in Patients With Deep Sternal Wound Infection.}},
  url          = {{https://lup.lub.lu.se/search/files/5270938/4731439}},
  volume       = {{14}},
  year         = {{2014}},
}