Submicrometer-resolution photoelectron-spectroscopy at MAX Lab
(1989) In Review of Scientific Instruments 60(7). p.2168-2171- Abstract
The design of a photoemission microprobe beamline at the MAX laboratory is presented. The beamline will utilize the radiation from a 26 period undulator on the 550-MeV MAX storage ring. The beamline will be capable of providing about 1010 photons/s of monochromatized radiation (bandwidth better than 0.25 eV) in the energy range of 20 to 150 eV focused into a submicrometer focal spot. It will be used for scanning photoemission experiments with a lateral resolution better than 1 μm. The optical system is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a specially designed ellipsoidal focusing mirror.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/b1db18b2-0346-48a0-b873-99cd5e871540
- author
- Nyholm, R. LU ; Eriksson, M. LU ; Hansen, K. LU ; Sairanen, O. P. ; Werin, S. LU ; Flodström, A. ; Törnevik, C. ; Meinander, T. and Sarakontu, M.
- organization
- publishing date
- 1989
- type
- Contribution to journal
- publication status
- published
- in
- Review of Scientific Instruments
- volume
- 60
- issue
- 7
- pages
- 4 pages
- publisher
- American Institute of Physics (AIP)
- external identifiers
-
- scopus:0042066916
- ISSN
- 0034-6748
- DOI
- 10.1063/1.1140812
- language
- English
- LU publication?
- yes
- id
- b1db18b2-0346-48a0-b873-99cd5e871540
- date added to LUP
- 2016-08-19 11:04:01
- date last changed
- 2021-12-29 08:03:34
@article{b1db18b2-0346-48a0-b873-99cd5e871540, abstract = {{<p>The design of a photoemission microprobe beamline at the MAX laboratory is presented. The beamline will utilize the radiation from a 26 period undulator on the 550-MeV MAX storage ring. The beamline will be capable of providing about 10<sup>10</sup> photons/s of monochromatized radiation (bandwidth better than 0.25 eV) in the energy range of 20 to 150 eV focused into a submicrometer focal spot. It will be used for scanning photoemission experiments with a lateral resolution better than 1 μm. The optical system is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a specially designed ellipsoidal focusing mirror.</p>}}, author = {{Nyholm, R. and Eriksson, M. and Hansen, K. and Sairanen, O. P. and Werin, S. and Flodström, A. and Törnevik, C. and Meinander, T. and Sarakontu, M.}}, issn = {{0034-6748}}, language = {{eng}}, number = {{7}}, pages = {{2168--2171}}, publisher = {{American Institute of Physics (AIP)}}, series = {{Review of Scientific Instruments}}, title = {{Submicrometer-resolution photoelectron-spectroscopy at MAX Lab}}, url = {{http://dx.doi.org/10.1063/1.1140812}}, doi = {{10.1063/1.1140812}}, volume = {{60}}, year = {{1989}}, }