The vacuum ultraviolet scanning photoelectron microscope at MAX-LAB
(1995) In Review of Scientific Instruments 66(2). p.1398-1400- Abstract
The performance of a scanning photoelectron microscope at the MAX I storage ring in Lund is presented. The microscope utilizes undulator radiation in the energy range 15-150 eV and is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a gracing incidence ellipsoidal focusing mirror. The instrument with its high photon flux, 109-1010 photons/s, and narrow bandwidth, better than 0.2 eV, is excellently suited for high-resolution core-level spectroscopy and imaging of samples with lateral inhomogeneties in the micrometer range.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/cda8af80-a90b-45c5-818e-a40e3e3b0411
- author
- Johansson, U. LU ; Nyholm, R. LU ; Törnevik, C. and Flodström, A.
- organization
- publishing date
- 1995
- type
- Contribution to journal
- publication status
- published
- in
- Review of Scientific Instruments
- volume
- 66
- issue
- 2
- pages
- 3 pages
- publisher
- American Institute of Physics (AIP)
- external identifiers
-
- scopus:0001414102
- ISSN
- 0034-6748
- DOI
- 10.1063/1.1145986
- language
- English
- LU publication?
- yes
- id
- cda8af80-a90b-45c5-818e-a40e3e3b0411
- date added to LUP
- 2016-04-29 11:01:53
- date last changed
- 2021-01-03 03:22:11
@article{cda8af80-a90b-45c5-818e-a40e3e3b0411, abstract = {{<p>The performance of a scanning photoelectron microscope at the MAX I storage ring in Lund is presented. The microscope utilizes undulator radiation in the energy range 15-150 eV and is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a gracing incidence ellipsoidal focusing mirror. The instrument with its high photon flux, 10<sup>9</sup>-10<sup>10</sup> photons/s, and narrow bandwidth, better than 0.2 eV, is excellently suited for high-resolution core-level spectroscopy and imaging of samples with lateral inhomogeneties in the micrometer range.</p>}}, author = {{Johansson, U. and Nyholm, R. and Törnevik, C. and Flodström, A.}}, issn = {{0034-6748}}, language = {{eng}}, number = {{2}}, pages = {{1398--1400}}, publisher = {{American Institute of Physics (AIP)}}, series = {{Review of Scientific Instruments}}, title = {{The vacuum ultraviolet scanning photoelectron microscope at MAX-LAB}}, url = {{http://dx.doi.org/10.1063/1.1145986}}, doi = {{10.1063/1.1145986}}, volume = {{66}}, year = {{1995}}, }