A new general-purpose isothermal microcalorimeter for use at temperatures up to 200 degrees C
(2003) In Thermochimica Acta 405(1). p.15-20- Abstract
- The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 degreesC. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use
f an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the... (More) - The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 degreesC. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use
f an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the temperature range 15-200 degreesC. Typical values obtained during 10 h periods at 200 degreesC are +/-3 and +/-10 nW for the baseline drift and baseline fluctuations, respectively. The heat detection limit, determined by release of electrical energy, is about 2 muJ. Preliminary stability measurements have been conducted at 100 degreesC on samples of stabilised and non-stabilised polyamide film. (C) 2003 Elsevier Science B.V. All rights reserved. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/122254
- author
- Wadsö, Ingemar LU and Wadsö, Lars LU
- organization
- publishing date
- 2003
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- Isothermal microcalorimeter, Nanowatt, 200 °C, Stability measurements
- in
- Thermochimica Acta
- volume
- 405
- issue
- 1
- pages
- 15 - 20
- publisher
- Elsevier
- external identifiers
-
- wos:000186127000003
- scopus:0344033884
- ISSN
- 0040-6031
- DOI
- 10.1016/S0040-6031(03)00124-2
- language
- English
- LU publication?
- yes
- id
- 59bcde93-ba0b-454a-b698-b8532ec3cc32 (old id 122254)
- date added to LUP
- 2016-04-01 15:25:55
- date last changed
- 2022-01-28 05:16:16
@article{59bcde93-ba0b-454a-b698-b8532ec3cc32, abstract = {{The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 degreesC. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use <br/><br> f an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the temperature range 15-200 degreesC. Typical values obtained during 10 h periods at 200 degreesC are +/-3 and +/-10 nW for the baseline drift and baseline fluctuations, respectively. The heat detection limit, determined by release of electrical energy, is about 2 muJ. Preliminary stability measurements have been conducted at 100 degreesC on samples of stabilised and non-stabilised polyamide film. (C) 2003 Elsevier Science B.V. All rights reserved.}}, author = {{Wadsö, Ingemar and Wadsö, Lars}}, issn = {{0040-6031}}, keywords = {{Isothermal microcalorimeter; Nanowatt; 200 °C; Stability measurements}}, language = {{eng}}, number = {{1}}, pages = {{15--20}}, publisher = {{Elsevier}}, series = {{Thermochimica Acta}}, title = {{A new general-purpose isothermal microcalorimeter for use at temperatures up to 200 degrees C}}, url = {{http://dx.doi.org/10.1016/S0040-6031(03)00124-2}}, doi = {{10.1016/S0040-6031(03)00124-2}}, volume = {{405}}, year = {{2003}}, }