Monte Carlo model of the Studsvik BNCT clinical beam : Description an validation

Giusti, Valerio ; Munck Af Rosenschöld, Per M. LU orcid ; Sköld, Kurt ; Montagnini, Bruno and Capala, Jacek (2003) In Medical Physics 30(12). p.3107-3117
Abstract

The neutron beam at the Studsvik facility for boron neutron capture therapy (BNCT) and the validation of the related computational model developed for the MCNP-4B Monte Carlo code are presented. Several measurements performed at the epithermal neutron port used for clinical trials have been made in order to validate the Monte Carlo computational model. The good general agreement between the MCNP calculations and the experimental results has provided an adequate check of the calculation procedure. In particular, at the nominal reactor power of 1 MW, the calculated in-air epithermal neutron flux in the energy interval between 0.4 eV-10 keV is 3.24 × 109 n cm -2 s-1 (±1.2% 1 std. dev.) while the measured... (More)

The neutron beam at the Studsvik facility for boron neutron capture therapy (BNCT) and the validation of the related computational model developed for the MCNP-4B Monte Carlo code are presented. Several measurements performed at the epithermal neutron port used for clinical trials have been made in order to validate the Monte Carlo computational model. The good general agreement between the MCNP calculations and the experimental results has provided an adequate check of the calculation procedure. In particular, at the nominal reactor power of 1 MW, the calculated in-air epithermal neutron flux in the energy interval between 0.4 eV-10 keV is 3.24 × 109 n cm -2 s-1 (±1.2% 1 std. dev.) while the measured value is 3.30 × 109 n cm-2 s-1 (±5.0% 1 std. dev.). Furthermore, the calculated in-phantom thermal neutron flux, equal to 6.43 × 109 n cm-2 s -1 (±1.0% 1 std. dev.), and the corresponding measured value of 6.33 × 109 n cm-2 s-1 (±5.3% 1 std. dev.) agree within their respective uncertainties. The only statistically significant disagreement is a discrepancy of 39% between the MCNP calculations of the in-air photon kerma and the corresponding experimental value. Despite this, a quite acceptable overall in-phantom beam performance was obtained, with a maximum value of the therapeutic ratio (the ratio between the local tumor dose and the maximum healthy tissue dose) equal to 6.7. The described MCNP model of the Studsvik facility has been deemed adequate to evaluate further improvements in the beam design as well as to plan experimental work.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
publishing date
type
Contribution to journal
publication status
published
subject
keywords
epithermal neutron beam, BNCT, nuclear reactor, Monte Carlo
in
Medical Physics
volume
30
issue
12
pages
3107 - 3117
publisher
American Association of Physicists in Medicine
external identifiers
  • wos:000187456800008
  • pmid:14713077
  • scopus:0346994903
  • pmid:14713077
ISSN
0094-2405
DOI
10.1118/1.1626120
language
English
LU publication?
no
id
04a3a21b-5dd1-4cf9-9ad5-a4e810400831 (old id 291308)
date added to LUP
2016-04-01 15:31:59
date last changed
2023-07-21 03:07:59
@article{04a3a21b-5dd1-4cf9-9ad5-a4e810400831,
  abstract     = {{<p>The neutron beam at the Studsvik facility for boron neutron capture therapy (BNCT) and the validation of the related computational model developed for the MCNP-4B Monte Carlo code are presented. Several measurements performed at the epithermal neutron port used for clinical trials have been made in order to validate the Monte Carlo computational model. The good general agreement between the MCNP calculations and the experimental results has provided an adequate check of the calculation procedure. In particular, at the nominal reactor power of 1 MW, the calculated in-air epithermal neutron flux in the energy interval between 0.4 eV-10 keV is 3.24 × 10<sup>9</sup> n cm <sup>-2</sup> s<sup>-1</sup> (±1.2% 1 std. dev.) while the measured value is 3.30 × 10<sup>9</sup> n cm<sup>-2</sup> s<sup>-1</sup> (±5.0% 1 std. dev.). Furthermore, the calculated in-phantom thermal neutron flux, equal to 6.43 × 10<sup>9</sup> n cm<sup>-2</sup> s <sup>-1</sup> (±1.0% 1 std. dev.), and the corresponding measured value of 6.33 × 10<sup>9</sup> n cm<sup>-2</sup> s<sup>-1</sup> (±5.3% 1 std. dev.) agree within their respective uncertainties. The only statistically significant disagreement is a discrepancy of 39% between the MCNP calculations of the in-air photon kerma and the corresponding experimental value. Despite this, a quite acceptable overall in-phantom beam performance was obtained, with a maximum value of the therapeutic ratio (the ratio between the local tumor dose and the maximum healthy tissue dose) equal to 6.7. The described MCNP model of the Studsvik facility has been deemed adequate to evaluate further improvements in the beam design as well as to plan experimental work.</p>}},
  author       = {{Giusti, Valerio and Munck Af Rosenschöld, Per M. and Sköld, Kurt and Montagnini, Bruno and Capala, Jacek}},
  issn         = {{0094-2405}},
  keywords     = {{epithermal neutron beam; BNCT; nuclear reactor; Monte Carlo}},
  language     = {{eng}},
  number       = {{12}},
  pages        = {{3107--3117}},
  publisher    = {{American Association of Physicists in Medicine}},
  series       = {{Medical Physics}},
  title        = {{Monte Carlo model of the Studsvik BNCT clinical beam : Description an validation}},
  url          = {{http://dx.doi.org/10.1118/1.1626120}},
  doi          = {{10.1118/1.1626120}},
  volume       = {{30}},
  year         = {{2003}},
}