European quadricentric evaluation of a breast MR biopsy and localization device: technical improvements based on phase-I evaluation

Prat, X ; Sittek, H ; Grosse, A ; Bååth, Lars LU ; Perlet, C ; Alberich, T ; Lamarque, JM ; Andersson, Ingvar LU ; Reiser, M and Taourel, P , et al. (2002) In European Radiology 12(7). p.1720-1727
Abstract
Our purpose was to report about technical success, problems and solutions, as experienced in a first multicentre study on MR-guided localisation or vacuum biopsy of breast lesions. The study was carried out at four European sites using a dedicated prototype breast biopsy device. Experiences with 49 scheduled localisation procedures and 188 vacuum biopsies are reported. Apart from 35 dropped indications, one localisation procedure and 9 vacuum biopsies were not possible (3 times space problems due to obesity, 2 times too strong compression, 3 times impaired access froth medially, 2 times impaired access due to a metal bar). Problems due to too strong compression were recognised by repeat MR without compression. During the procedure problems... (More)
Our purpose was to report about technical success, problems and solutions, as experienced in a first multicentre study on MR-guided localisation or vacuum biopsy of breast lesions. The study was carried out at four European sites using a dedicated prototype breast biopsy device. Experiences with 49 scheduled localisation procedures and 188 vacuum biopsies are reported. Apart from 35 dropped indications, one localisation procedure and 9 vacuum biopsies were not possible (3 times space problems due to obesity, 2 times too strong compression, 3 times impaired access froth medially, 2 times impaired access due to a metal bar). Problems due to too strong compression were recognised by repeat MR without compression. During the procedure problems leading to an uncertain result occurred in eight vacuum biopsies, two related to the procedure: one limited access, and one strong post-biopsy enhancement. Improvements after phase-I study concerned removal of the metal bar, development of an improved medial access, of a profile imitating the biopsy gun, optimisation of compression plates and improved software support. The partners agreed that the improvements answered all important technical problems. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and , et al. (More)
; ; ; ; ; ; ; ; ; ; and (Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
breast biopsy, breast MR
in
European Radiology
volume
12
issue
7
pages
1720 - 1727
publisher
Springer
external identifiers
  • wos:000177003700013
  • pmid:12111063
  • scopus:0036634119
ISSN
0938-7994
DOI
10.1007/s00330-002-1317-2
language
English
LU publication?
yes
id
6a6af119-8365-4e61-8b63-318fc75ebb71 (old id 332752)
date added to LUP
2016-04-01 11:34:44
date last changed
2022-01-26 07:15:06
@article{6a6af119-8365-4e61-8b63-318fc75ebb71,
  abstract     = {{Our purpose was to report about technical success, problems and solutions, as experienced in a first multicentre study on MR-guided localisation or vacuum biopsy of breast lesions. The study was carried out at four European sites using a dedicated prototype breast biopsy device. Experiences with 49 scheduled localisation procedures and 188 vacuum biopsies are reported. Apart from 35 dropped indications, one localisation procedure and 9 vacuum biopsies were not possible (3 times space problems due to obesity, 2 times too strong compression, 3 times impaired access froth medially, 2 times impaired access due to a metal bar). Problems due to too strong compression were recognised by repeat MR without compression. During the procedure problems leading to an uncertain result occurred in eight vacuum biopsies, two related to the procedure: one limited access, and one strong post-biopsy enhancement. Improvements after phase-I study concerned removal of the metal bar, development of an improved medial access, of a profile imitating the biopsy gun, optimisation of compression plates and improved software support. The partners agreed that the improvements answered all important technical problems.}},
  author       = {{Prat, X and Sittek, H and Grosse, A and Bååth, Lars and Perlet, C and Alberich, T and Lamarque, JM and Andersson, Ingvar and Reiser, M and Taourel, P and Fischer, H and Heywang-Kobrunner, SH}},
  issn         = {{0938-7994}},
  keywords     = {{breast biopsy; breast MR}},
  language     = {{eng}},
  number       = {{7}},
  pages        = {{1720--1727}},
  publisher    = {{Springer}},
  series       = {{European Radiology}},
  title        = {{European quadricentric evaluation of a breast MR biopsy and localization device: technical improvements based on phase-I evaluation}},
  url          = {{http://dx.doi.org/10.1007/s00330-002-1317-2}},
  doi          = {{10.1007/s00330-002-1317-2}},
  volume       = {{12}},
  year         = {{2002}},
}