A Microdialysis probe coupled with a miniaturized thermal glucose sensor for in vivo monitoring

Amine, A. ; Digua, K. ; Xie, B. LU and Danielsson, B. LU (1995) In Analytical Letters 28(13). p.2275-2286
Abstract

A miniaturized thermal flow injection analysis biosensor has been
coupled with a microdialysis probe for continuous subcutaneous glucose
monitoring. Thermal biosensors are based on the principle of measuring
the heat evolved during enzyme catalysed reactions. The system presented
here consists of a miniaturized thermal biosensor with a small column
containing coimmibolized glucose oxidase and catalase. The analysis
buffer passes through the column at a flow rate of 60μL/min via an 1μL
sample loop which is connected to a microdialysis probe.

Invitro
results showed constant permeability of the probe and stability of the
biosensor response during 24 hours. The response time was 85 sec... (More)

A miniaturized thermal flow injection analysis biosensor has been
coupled with a microdialysis probe for continuous subcutaneous glucose
monitoring. Thermal biosensors are based on the principle of measuring
the heat evolved during enzyme catalysed reactions. The system presented
here consists of a miniaturized thermal biosensor with a small column
containing coimmibolized glucose oxidase and catalase. The analysis
buffer passes through the column at a flow rate of 60μL/min via an 1μL
sample loop which is connected to a microdialysis probe.

Invitro
results showed constant permeability of the probe and stability of the
biosensor response during 24 hours. The response time was 85 sec giving a
sample rate of 42 samples/hour.

During a load experiment, the
glucose profile in a healthy volunteer was followed both in the
subcutaneous tissue and blood using the microdialysis set-up proposed
and comparing to blood glucose analyser.

(Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Glucose Biosensor, In-Vivo Monitoring, Microdialysis, Thermal
in
Analytical Letters
volume
28
issue
13
pages
12 pages
publisher
Taylor & Francis
external identifiers
  • scopus:0028814787
ISSN
0003-2719
DOI
10.1080/00032719508000371
language
English
LU publication?
yes
id
363ffe68-a159-4ee5-93d3-d6c4a95f610d
date added to LUP
2021-10-27 15:51:56
date last changed
2021-11-01 11:36:16
@article{363ffe68-a159-4ee5-93d3-d6c4a95f610d,
  abstract     = {{<p>A miniaturized thermal flow injection analysis biosensor has been <br>
coupled with a microdialysis probe for continuous subcutaneous glucose <br>
monitoring. Thermal biosensors are based on the principle of measuring <br>
the heat evolved during enzyme catalysed reactions. The system presented<br>
 here consists of a miniaturized thermal biosensor with a small column <br>
containing coimmibolized glucose oxidase and catalase. The analysis <br>
buffer passes through the column at a flow rate of 60μL/min via an 1μL <br>
sample loop which is connected to a microdialysis probe.</p> <p>Invitro <br>
results showed constant permeability of the probe and stability of the <br>
biosensor response during 24 hours. The response time was 85 sec giving a<br>
 sample rate of 42 samples/hour.</p> <p>During a load experiment, the <br>
glucose profile in a healthy volunteer was followed both in the <br>
subcutaneous tissue and blood using the microdialysis set-up proposed <br>
and comparing to blood glucose analyser.</p>}},
  author       = {{Amine, A. and Digua, K. and Xie, B. and Danielsson, B.}},
  issn         = {{0003-2719}},
  keywords     = {{Glucose Biosensor; In-Vivo Monitoring; Microdialysis; Thermal}},
  language     = {{eng}},
  number       = {{13}},
  pages        = {{2275--2286}},
  publisher    = {{Taylor & Francis}},
  series       = {{Analytical Letters}},
  title        = {{A Microdialysis probe coupled with a miniaturized thermal glucose sensor for in vivo monitoring}},
  url          = {{http://dx.doi.org/10.1080/00032719508000371}},
  doi          = {{10.1080/00032719508000371}},
  volume       = {{28}},
  year         = {{1995}},
}