Abstract
Ultrafast optical spectroscopy techniques are often employed to gain information about samples that are liquid at room temperature and frozen at cryogenic temperatures. However, the measurements suffer from the presence of unwanted, non-resonant signals originating in the sample cell walls. Most of these artifacts can be avoided in the measurements performed at room temperature by using liquid jet systems, i.e., by removing the sample cell. However, these systems cannot be used in low temperature measurements, when the sample is frozen. Herein we describe a freestanding sample holder that allows low temperature ultrafast spectroscopy measurements free of artifacts caused by the sample cell.
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Review of Scientific Instruments
volume
85
issue
4
article number
043103
publisher
American Institute of Physics (AIP)
external identifiers
  • pmid:24784589
  • wos:000335920600003
  • scopus:84897996474
  • pmid:24784589
ISSN
1089-7623
DOI
10.1063/1.4870277
language
English
LU publication?
yes
additional info
The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Chemical Physics (S) (011001060)
id
2111640b-a66b-493d-bde4-7be501fb856d (old id 4456067)
date added to LUP
2016-04-01 10:18:30
date last changed
2023-08-30 23:28:24
@article{2111640b-a66b-493d-bde4-7be501fb856d,
  abstract     = {{Ultrafast optical spectroscopy techniques are often employed to gain information about samples that are liquid at room temperature and frozen at cryogenic temperatures. However, the measurements suffer from the presence of unwanted, non-resonant signals originating in the sample cell walls. Most of these artifacts can be avoided in the measurements performed at room temperature by using liquid jet systems, i.e., by removing the sample cell. However, these systems cannot be used in low temperature measurements, when the sample is frozen. Herein we describe a freestanding sample holder that allows low temperature ultrafast spectroscopy measurements free of artifacts caused by the sample cell.}},
  author       = {{Yoo, Sunny and Alster, Jan and Zigmantas, Donatas}},
  issn         = {{1089-7623}},
  language     = {{eng}},
  number       = {{4}},
  publisher    = {{American Institute of Physics (AIP)}},
  series       = {{Review of Scientific Instruments}},
  title        = {{Freestanding sample holder for ultrafast optical spectroscopy at low temperatures.}},
  url          = {{https://lup.lub.lu.se/search/files/102846815/Yoo2014RSI_freestanding_sample_holder.pdf}},
  doi          = {{10.1063/1.4870277}},
  volume       = {{85}},
  year         = {{2014}},
}