OPTIMIZATION OF PHOTOEMISSION BLADE X-RAY BEAM POSITION MONITOR FOR WIDE RANGE OF UNDULATOR GAP OPERATION

Ilinski, P. LU (2025) 14th International Beam Instrumentation Conference, IBIC 2025 In Proceedings of the International Beam Instrumentation Conference, IBIC p.781-783
Abstract

Optimization of the photoemission blade type XBPM mostly performed to increase its resolution by increasing the readout signal current and by optimizing blades geometry. This requires tailoring of the XBPM for a particular undulator, making almost every XBPM unique at the Synchrotron Radiation facility. In many cases the calibration coefficient is gap dependent. To overcome those drawbacks an alternative optimization strategy is considered which will help to minimize gap dependence and could provide a unified XBPM design for all XBPM at Synchrotron Radiation (SR) facility.

Please use this url to cite or link to this publication:
author
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
host publication
Proceedings of the 14th International Beam Instrumentation Conference
series title
Proceedings of the International Beam Instrumentation Conference, IBIC
editor
Kumar, Narender
pages
3 pages
publisher
JACoW Publishing
conference name
14th International Beam Instrumentation Conference, IBIC 2025
conference location
Liverpool, United Kingdom
conference dates
2025-09-07 - 2025-09-11
external identifiers
  • scopus:105038456670
ISSN
2673-5350
ISBN
9783954502622
DOI
10.18429/JACoW-IBIC2025-WEPMO24
language
English
LU publication?
yes
additional info
Publisher Copyright: © 2025 International Beam Instrumentation Conference.All rights reserved.
id
76ad1af7-23bb-41fd-9e48-85f689639e99
date added to LUP
2026-07-16 14:13:45
date last changed
2026-08-10 08:07:15
@inproceedings{76ad1af7-23bb-41fd-9e48-85f689639e99,
  abstract     = {{<p>Optimization of the photoemission blade type XBPM mostly performed to increase its resolution by increasing the readout signal current and by optimizing blades geometry. This requires tailoring of the XBPM for a particular undulator, making almost every XBPM unique at the Synchrotron Radiation facility. In many cases the calibration coefficient is gap dependent. To overcome those drawbacks an alternative optimization strategy is considered which will help to minimize gap dependence and could provide a unified XBPM design for all XBPM at Synchrotron Radiation (SR) facility.</p>}},
  author       = {{Ilinski, P.}},
  booktitle    = {{Proceedings of the 14th International Beam Instrumentation Conference}},
  editor       = {{Kumar, Narender}},
  isbn         = {{9783954502622}},
  issn         = {{2673-5350}},
  language     = {{eng}},
  pages        = {{781--783}},
  publisher    = {{JACoW Publishing}},