Characterization of the response function of a Si(Li) detector using an absorber technique

Larsson, N. P.O. LU ; Tapper, U. A.S. and Martinsson, B. G. LU (1989) In Nuclear Inst. and Methods in Physics Research, B 43(4). p.574-580
Abstract

The non-Gaussian response function of a Si(Li) detector has been measured for characteristic Kα X-rays in the energy interval 1.5-8.6 keV using an absorber technique. The method is based on the different response to an attenuating absorber placed in front of the detector for the detector tail as compared to fully absorbed photons in the same position in the spectrum. The shape of the tail of a Kα X-ray peak could be derived from three X-ray spectra acquired with different attenuating absorbers using the PIXE or XRF method for excitation of single element standards. The line shape function could be described by a full energy Gaussian, a shelf, an exponential and a Si-escape component. The results presented are in... (More)

The non-Gaussian response function of a Si(Li) detector has been measured for characteristic Kα X-rays in the energy interval 1.5-8.6 keV using an absorber technique. The method is based on the different response to an attenuating absorber placed in front of the detector for the detector tail as compared to fully absorbed photons in the same position in the spectrum. The shape of the tail of a Kα X-ray peak could be derived from three X-ray spectra acquired with different attenuating absorbers using the PIXE or XRF method for excitation of single element standards. The line shape function could be described by a full energy Gaussian, a shelf, an exponential and a Si-escape component. The results presented are in good agreement with previously reported response functions obtained in measurements of monochromatic X-rays. The results will be implemented in a computer program for evaluating micro-PIXE spectra.

(Less)
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Nuclear Inst. and Methods in Physics Research, B
volume
43
issue
4
pages
7 pages
publisher
Elsevier
external identifiers
  • scopus:0024754750
ISSN
0168-583X
DOI
10.1016/0168-583X(89)90409-6
language
English
LU publication?
yes
id
77505cba-a637-41b0-8757-6415ea246f6c
date added to LUP
2019-05-16 09:48:35
date last changed
2021-01-03 10:28:56
@article{77505cba-a637-41b0-8757-6415ea246f6c,
  abstract     = {{<p>The non-Gaussian response function of a Si(Li) detector has been measured for characteristic K<sub>α</sub> X-rays in the energy interval 1.5-8.6 keV using an absorber technique. The method is based on the different response to an attenuating absorber placed in front of the detector for the detector tail as compared to fully absorbed photons in the same position in the spectrum. The shape of the tail of a K<sub>α</sub> X-ray peak could be derived from three X-ray spectra acquired with different attenuating absorbers using the PIXE or XRF method for excitation of single element standards. The line shape function could be described by a full energy Gaussian, a shelf, an exponential and a Si-escape component. The results presented are in good agreement with previously reported response functions obtained in measurements of monochromatic X-rays. The results will be implemented in a computer program for evaluating micro-PIXE spectra.</p>}},
  author       = {{Larsson, N. P.O. and Tapper, U. A.S. and Martinsson, B. G.}},
  issn         = {{0168-583X}},
  language     = {{eng}},
  number       = {{4}},
  pages        = {{574--580}},
  publisher    = {{Elsevier}},
  series       = {{Nuclear Inst. and Methods in Physics Research, B}},
  title        = {{Characterization of the response function of a Si(Li) detector using an absorber technique}},
  url          = {{http://dx.doi.org/10.1016/0168-583X(89)90409-6}},
  doi          = {{10.1016/0168-583X(89)90409-6}},
  volume       = {{43}},
  year         = {{1989}},
}