Can scatter correction software replace a grid in DR pelvic examinations?

Precht, Helle ; Mørup, Svea Deppe ; Tingberg, Anders LU orcid ; Outzen, Claus Bjørn ; Kusk, Kirsten Weber ; Nielsen, Rita Møller ; Midtgård, Marie ; Winther, Martin Brændegaard ; Waaler, Dag and Kusk, Martin Weber (2019) In Radiation Protection Dosimetry 187(1). p.8-16
Abstract

The purpose was to examine if scatter correction software could replace a grid while maintaining image quality and reducing radiation dose for pelvic DR examinations. Grid images was produced with 70 kV and 16mAs. Anthropomorphic- and Contrast Detail RADiography (CDRAD) non-grid images were produced with 60 kV, 80 kV and 90 kV combined with five different mAs and scatter correction software. The anthropomorphic images were analyzed by absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed using the CDRAD analysis software. The results showed a total of 54.6% non-grid images were evaluated as unsuitable for diagnostic use by the VGA. The CDRAD grid images showed that the IQF_inv values were significantly different (p =... (More)

The purpose was to examine if scatter correction software could replace a grid while maintaining image quality and reducing radiation dose for pelvic DR examinations. Grid images was produced with 70 kV and 16mAs. Anthropomorphic- and Contrast Detail RADiography (CDRAD) non-grid images were produced with 60 kV, 80 kV and 90 kV combined with five different mAs and scatter correction software. The anthropomorphic images were analyzed by absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed using the CDRAD analysis software. The results showed a total of 54.6% non-grid images were evaluated as unsuitable for diagnostic use by the VGA. The CDRAD grid images showed that the IQF_inv values were significantly different (p = 0.0001) when compared to every group of non-grid images. Hereby, the conclusion stated that the scatter correction software did not compensate for the loss in image quality due to scattered radiation at the exposure levels included in a pelvic examination.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Digital radiography, image quality, pelvic, Radiation dose, Scatter correction software
in
Radiation Protection Dosimetry
volume
187
issue
1
pages
9 pages
publisher
Oxford University Press
external identifiers
  • pmid:31111927
  • scopus:85084961412
ISSN
1742-3406
DOI
10.1093/rpd/ncz129
language
English
LU publication?
yes
id
79db9705-dbbc-4bd6-b839-1ed1349eedba
date added to LUP
2020-06-23 13:12:21
date last changed
2024-10-03 04:24:18
@article{79db9705-dbbc-4bd6-b839-1ed1349eedba,
  abstract     = {{<p>The purpose was to examine if scatter correction software could replace a grid while maintaining image quality and reducing radiation dose for pelvic DR examinations. Grid images was produced with 70 kV and 16mAs. Anthropomorphic- and Contrast Detail RADiography (CDRAD) non-grid images were produced with 60 kV, 80 kV and 90 kV combined with five different mAs and scatter correction software. The anthropomorphic images were analyzed by absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed using the CDRAD analysis software. The results showed a total of 54.6% non-grid images were evaluated as unsuitable for diagnostic use by the VGA. The CDRAD grid images showed that the IQF_inv values were significantly different (p = 0.0001) when compared to every group of non-grid images. Hereby, the conclusion stated that the scatter correction software did not compensate for the loss in image quality due to scattered radiation at the exposure levels included in a pelvic examination.</p>}},
  author       = {{Precht, Helle and Mørup, Svea Deppe and Tingberg, Anders and Outzen, Claus Bjørn and Kusk, Kirsten Weber and Nielsen, Rita Møller and Midtgård, Marie and Winther, Martin Brændegaard and Waaler, Dag and Kusk, Martin Weber}},
  issn         = {{1742-3406}},
  keywords     = {{Digital radiography; image quality; pelvic; Radiation dose; Scatter correction software}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{8--16}},
  publisher    = {{Oxford University Press}},
  series       = {{Radiation Protection Dosimetry}},
  title        = {{Can scatter correction software replace a grid in DR pelvic examinations?}},
  url          = {{http://dx.doi.org/10.1093/rpd/ncz129}},
  doi          = {{10.1093/rpd/ncz129}},
  volume       = {{187}},
  year         = {{2019}},
}