Multicenter study for the evaluation of a dedicated biopsy device for MR-guided vacuum biopsy of the breast

Perlet, C ; Heinig, A ; Prat, X ; Casselman, J ; Bååth, Lars LU ; Sittek, H ; Stets, C ; Lamarque, J ; Anderson, I and Schneider, P , et al. (2002) In European Radiology 12(6). p.1463-1470
Abstract
The purpose of this multicenter study was to determine the accuracy and clinical value of a dedicated breast biopsy system which allows for MR-guided vacuum biopsy (VB) of contrast-enhancing lesions. In five European centers, MR-guided 11-gauge VB was performed on 341 lesions. In 7 cases VB was unsuccessful. This was immediately realized on postinterventional images or direct follow-up combined with histopathology-imaging correlation; thus, a false-negative diagnosis was avoided. Histology of 334 successful biopsies yielded 84 (25%) malignancies, 17 (5%) atypical ductal hyperplasias, and 233 (70%) benign entities. Verification of malignant or borderline lesions included re-excision of the biopsy cavity. Benign histologic biopsy results... (More)
The purpose of this multicenter study was to determine the accuracy and clinical value of a dedicated breast biopsy system which allows for MR-guided vacuum biopsy (VB) of contrast-enhancing lesions. In five European centers, MR-guided 11-gauge VB was performed on 341 lesions. In 7 cases VB was unsuccessful. This was immediately realized on postinterventional images or direct follow-up combined with histopathology-imaging correlation; thus, a false-negative diagnosis was avoided. Histology of 334 successful biopsies yielded 84 (25%) malignancies, 17 (5%) atypical ductal hyperplasias, and 233 (70%) benign entities. Verification of malignant or borderline lesions included re-excision of the biopsy cavity. Benign histologic biopsy results were verified by retrospective correlation with the pre- and postinterventional MRI and by subsequent follow-up. Our results indicate that MR-guided VB, in combination with the dedicated biopsy coil, offers the possibility to accurately diagnose even very small lesions that can only be visualized or localized by MRI. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and , et al. (More)
; ; ; ; ; ; ; ; ; ; ; and (Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
intervention, biopsy, breast, MR imaging
in
European Radiology
volume
12
issue
6
pages
1463 - 1470
publisher
Springer
external identifiers
  • wos:000176197400028
  • pmid:12042955
  • scopus:0036598381
  • pmid:12042955
ISSN
0938-7994
DOI
10.1007/s00330-002-1376-4
language
English
LU publication?
yes
id
9effb180-9b50-406a-b09c-524ac780e5d8 (old id 893119)
date added to LUP
2016-04-01 12:22:12
date last changed
2022-01-27 02:48:41
@article{9effb180-9b50-406a-b09c-524ac780e5d8,
  abstract     = {{The purpose of this multicenter study was to determine the accuracy and clinical value of a dedicated breast biopsy system which allows for MR-guided vacuum biopsy (VB) of contrast-enhancing lesions. In five European centers, MR-guided 11-gauge VB was performed on 341 lesions. In 7 cases VB was unsuccessful. This was immediately realized on postinterventional images or direct follow-up combined with histopathology-imaging correlation; thus, a false-negative diagnosis was avoided. Histology of 334 successful biopsies yielded 84 (25%) malignancies, 17 (5%) atypical ductal hyperplasias, and 233 (70%) benign entities. Verification of malignant or borderline lesions included re-excision of the biopsy cavity. Benign histologic biopsy results were verified by retrospective correlation with the pre- and postinterventional MRI and by subsequent follow-up. Our results indicate that MR-guided VB, in combination with the dedicated biopsy coil, offers the possibility to accurately diagnose even very small lesions that can only be visualized or localized by MRI.}},
  author       = {{Perlet, C and Heinig, A and Prat, X and Casselman, J and Bååth, Lars and Sittek, H and Stets, C and Lamarque, J and Anderson, I and Schneider, P and Taourel, P and Reiser, M and Heywang-Kobrunner, SH}},
  issn         = {{0938-7994}},
  keywords     = {{intervention; biopsy; breast; MR imaging}},
  language     = {{eng}},
  number       = {{6}},
  pages        = {{1463--1470}},
  publisher    = {{Springer}},
  series       = {{European Radiology}},
  title        = {{Multicenter study for the evaluation of a dedicated biopsy device for MR-guided vacuum biopsy of the breast}},
  url          = {{http://dx.doi.org/10.1007/s00330-002-1376-4}},
  doi          = {{10.1007/s00330-002-1376-4}},
  volume       = {{12}},
  year         = {{2002}},
}