Quality Controls in Digital Mammography protocol of the EFOMP Mammo Working group

Gennaro, Gisella ; Avramova-Cholakova, Simona ; Azzalini, Andrea ; Luisa Chapel, Maria ; Chevalier, Margarita ; Ciraj, Olivera ; de las Heras, Hugo ; Gershan, Vesna ; Hemdal, Bengt LU and Keavey, Elizabeth , et al. (2018) In Physica Medica 48. p.55-64
Abstract

This article aims to present the protocol on Quality Controls in Digital Mammography published online in 2015 by the European Federation of Organisations for Medical Physics (EFOMP) which was developed by a Task Force under the Mammo Working Group. The main objective of this protocol was to define a minimum set of easily implemented quality control tests on digital mammography systems that can be used to assure the performance of a system within a set and acceptable range. Detailed step-by-step instructions have been provided, limiting as much as possible any misinterpretations or variations by the person performing. It is intended that these tests be implemented as part of the daily routine of medical physicists and system users... (More)

This article aims to present the protocol on Quality Controls in Digital Mammography published online in 2015 by the European Federation of Organisations for Medical Physics (EFOMP) which was developed by a Task Force under the Mammo Working Group. The main objective of this protocol was to define a minimum set of easily implemented quality control tests on digital mammography systems that can be used to assure the performance of a system within a set and acceptable range. Detailed step-by-step instructions have been provided, limiting as much as possible any misinterpretations or variations by the person performing. It is intended that these tests be implemented as part of the daily routine of medical physicists and system users throughout Europe in a harmonised way so allowing results to be compared. In this paper the main characteristics of the protocol are illustrated, including examples, together with a brief summary of the contents of each chapter. Finally, instructions for the download of the full protocol and of the related software tools are provided.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and , et al. (More)
; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; and (Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Digital mammography, EFOMP, Protocol, Quality control
in
Physica Medica
volume
48
pages
10 pages
publisher
ISTITUTI EDITORIALI E POLGRAFICI INTERNAZIONALI
external identifiers
  • scopus:85044929459
  • pmid:29728229
ISSN
1120-1797
DOI
10.1016/j.ejmp.2018.03.016
language
English
LU publication?
yes
id
9369bb13-a24e-428a-9691-9e7a3d1963f0
date added to LUP
2018-04-19 15:29:37
date last changed
2024-04-01 04:38:05
@article{9369bb13-a24e-428a-9691-9e7a3d1963f0,
  abstract     = {{<p>This article aims to present the protocol on Quality Controls in Digital Mammography published online in 2015 by the European Federation of Organisations for Medical Physics (EFOMP) which was developed by a Task Force under the Mammo Working Group. The main objective of this protocol was to define a minimum set of easily implemented quality control tests on digital mammography systems that can be used to assure the performance of a system within a set and acceptable range. Detailed step-by-step instructions have been provided, limiting as much as possible any misinterpretations or variations by the person performing. It is intended that these tests be implemented as part of the daily routine of medical physicists and system users throughout Europe in a harmonised way so allowing results to be compared. In this paper the main characteristics of the protocol are illustrated, including examples, together with a brief summary of the contents of each chapter. Finally, instructions for the download of the full protocol and of the related software tools are provided.</p>}},
  author       = {{Gennaro, Gisella and Avramova-Cholakova, Simona and Azzalini, Andrea and Luisa Chapel, Maria and Chevalier, Margarita and Ciraj, Olivera and de las Heras, Hugo and Gershan, Vesna and Hemdal, Bengt and Keavey, Elizabeth and Lanconelli, Nico and Menhart, Susanne and João Fartaria, Màrio and Pascoal, Ana and Pedersen, Kristin and Rivetti, Stefano and Rossetti, Veronica and Semturs, Friedrich and Sharp, Peter and Torresin, Alberto}},
  issn         = {{1120-1797}},
  keywords     = {{Digital mammography; EFOMP; Protocol; Quality control}},
  language     = {{eng}},
  month        = {{04}},
  pages        = {{55--64}},
  publisher    = {{ISTITUTI EDITORIALI E POLGRAFICI INTERNAZIONALI}},
  series       = {{Physica Medica}},
  title        = {{Quality Controls in Digital Mammography protocol of the EFOMP Mammo Working group}},
  url          = {{http://dx.doi.org/10.1016/j.ejmp.2018.03.016}},
  doi          = {{10.1016/j.ejmp.2018.03.016}},
  volume       = {{48}},
  year         = {{2018}},
}