Ceric Andelius, Irma LU ; Stenvall, Anna LU ; Nilsson, Elias ; Larsson, Erik LU and Gustafsson, Johan LU orcid (2026) In EJNMMI Physics 13(1).
Abstract

Background: The aim was to evaluate bias and precision for 177Lu-activity-quantification using window-based scatter compensation in a ring-configured CZT gamma camera (StarGuide™, GE HealthCare). This paper extends a previous study by applying modified scatter compensation (MSC) methods introduced with the StarGuidePlus upgrade. The modified methods account for contamination of scatter windows by primary photons due to tailing. Whilst the original compensation addressed tailing in the 208 keV peak only, the upgrade extends the concept to the 113 keV peak and enables the estimated primary signal in the scatter windows to be included to the measured projections (MSC + P). Methods: Listmode-data for a NEMA phantom and... (More)

Background: The aim was to evaluate bias and precision for 177Lu-activity-quantification using window-based scatter compensation in a ring-configured CZT gamma camera (StarGuide™, GE HealthCare). This paper extends a previous study by applying modified scatter compensation (MSC) methods introduced with the StarGuidePlus upgrade. The modified methods account for contamination of scatter windows by primary photons due to tailing. Whilst the original compensation addressed tailing in the 208 keV peak only, the upgrade extends the concept to the 113 keV peak and enables the estimated primary signal in the scatter windows to be included to the measured projections (MSC + P). Methods: Listmode-data for a NEMA phantom and anthropomorphic phantom were reframed to accommodate the updated settings for MSC and MSC + P. Reconstruction was performed using OS-EM (two to 30 iterations, 10 subsets) with compensation for attenuation, scatter (DEW/TEW, MSC and MSC + P), spatial resolution, and penetration at 208 keV. Volumes-of-interest following the manufacturer-specified sphere sizes were defined and activity-concentration was quantified for each sphere and the total phantom activity across six (10 min) time-frames. Bias was evaluated with mean relative error and precision with coefficient of variation (CV). Results: Imaging at 208 keV generally results in similar activity concentrations and total activity estimates for MSC and MSC + P (mean relative error: − 24.8 to − 23.4%, CV: 0.6 and 0.7% for the larges sphere). For 113 keV, a slightly better precision is obtained for MSC and MSC + P (mean relative error: − 15.4 and − 21.5%, CV: 0.6 and 0.7% for the largest sphere in the NEMA phantom) but a larger bias compared with TEW (mean relative error − 7.1%, CV 0.8%). The major difference is for total activity where the MSC and MSC + P decrease bias compared to TEW (5 and 4% versus 21% for 113 keV in the NEMA phantom). Conclusions: The improvement in activity-concentration accuracy in regard to bias and precision is modest in small-volume high activity-concentration regions. However, the findings suggest potential value in preservation of total activity for both MSC and MSC + P.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
CZT, Lutetium-177, Quantitative SPECT, Ring-configured SPECT, Scatter compensation
in
EJNMMI Physics
volume
13
issue
1
article number
65
publisher
Springer Science and Business Media B.V.
external identifiers
  • pmid:42289023
  • scopus:105041966599
ISSN
2197-7364
DOI
10.1186/s40658-026-00905-0
language
English
LU publication?
yes
additional info
Publisher Copyright: © The Author(s) 2026.
id
cefd4027-99e2-47b6-a3e0-2ff26c475f7f
date added to LUP
2026-07-23 10:04:48
date last changed
2026-09-03 13:19:37
@article{cefd4027-99e2-47b6-a3e0-2ff26c475f7f,
  abstract     = {{<p>Background: The aim was to evaluate bias and precision for <sup>177</sup>Lu-activity-quantification using window-based scatter compensation in a ring-configured CZT gamma camera (StarGuide™, GE HealthCare). This paper extends a previous study by applying modified scatter compensation (MSC) methods introduced with the StarGuidePlus upgrade. The modified methods account for contamination of scatter windows by primary photons due to tailing. Whilst the original compensation addressed tailing in the 208 keV peak only, the upgrade extends the concept to the 113 keV peak and enables the estimated primary signal in the scatter windows to be included to the measured projections (MSC + P). Methods: Listmode-data for a NEMA phantom and anthropomorphic phantom were reframed to accommodate the updated settings for MSC and MSC + P. Reconstruction was performed using OS-EM (two to 30 iterations, 10 subsets) with compensation for attenuation, scatter (DEW/TEW, MSC and MSC + P), spatial resolution, and penetration at 208 keV. Volumes-of-interest following the manufacturer-specified sphere sizes were defined and activity-concentration was quantified for each sphere and the total phantom activity across six (10 min) time-frames. Bias was evaluated with mean relative error and precision with coefficient of variation (CV). Results: Imaging at 208 keV generally results in similar activity concentrations and total activity estimates for MSC and MSC + P (mean relative error: − 24.8 to − 23.4%, CV: 0.6 and 0.7% for the larges sphere). For 113 keV, a slightly better precision is obtained for MSC and MSC + P (mean relative error: − 15.4 and − 21.5%, CV: 0.6 and 0.7% for the largest sphere in the NEMA phantom) but a larger bias compared with TEW (mean relative error − 7.1%, CV 0.8%). The major difference is for total activity where the MSC and MSC + P decrease bias compared to TEW (5 and 4% versus 21% for 113 keV in the NEMA phantom). Conclusions: The improvement in activity-concentration accuracy in regard to bias and precision is modest in small-volume high activity-concentration regions. However, the findings suggest potential value in preservation of total activity for both MSC and MSC + P.</p>}},
  author       = {{Ceric Andelius, Irma and Stenvall, Anna and Nilsson, Elias and Larsson, Erik and Gustafsson, Johan}},
  issn         = {{2197-7364}},
  keywords     = {{CZT; Lutetium-177; Quantitative SPECT; Ring-configured SPECT; Scatter compensation}},
  language     = {{eng}},
  number       = {{1}},
  publisher    = {{Springer Science and Business Media B.V.}},
  series       = {{EJNMMI Physics}},
  title        = {{Evaluation of modified window-based scatter compensation in quantitative <sup>177</sup>Lu-SPECT for a ring-configured CZT SPECT-CT}},
  url          = {{http://dx.doi.org/10.1186/s40658-026-00905-0}},
  doi          = {{10.1186/s40658-026-00905-0}},
  volume       = {{13}},
  year         = {{2026}},
}