Dose evaluation of simultaneous breast radiography and mechanical imaging
(2023) SPIE Medical Imaging 2023 12463.- Abstract
- This study investigates the impact in terms of radiation dose when performing simultaneous digital breast tomosynthesis
(DBT) and mechanical imaging (MI) – DBTMI. DBTMI has demonstrated the potential to increase specificity of cancer
detection, and reduce unnecessary biopsies, as compared to digital mammography (DM) screening. The presence of the
MI sensor during simultaneous image acquisition may increase the radiation dose when automatic exposure control is used.
In this project, a radiation dose study was conducted on clinically available breast imaging systems with and without the
MI sensor. We have investigated three approaches to analyse the dose increase in DBTMI, using (i) the estimates of average
glandular... (More) - This study investigates the impact in terms of radiation dose when performing simultaneous digital breast tomosynthesis
(DBT) and mechanical imaging (MI) – DBTMI. DBTMI has demonstrated the potential to increase specificity of cancer
detection, and reduce unnecessary biopsies, as compared to digital mammography (DM) screening. The presence of the
MI sensor during simultaneous image acquisition may increase the radiation dose when automatic exposure control is used.
In this project, a radiation dose study was conducted on clinically available breast imaging systems with and without the
MI sensor. We have investigated three approaches to analyse the dose increase in DBTMI, using (i) the estimates of average
glandular dose (AGD) reported in DICOM headers of radiography images; (ii) AGD measured by a conventional
dosemeter; and (iii) AGD measured by optically stimulated luminescence using NaCl pellets. The relative increase in AGD
estimated from DICOM headers when using the MI sensor was on average 10.7% and 12.4%, for DM and DBT
measurements, respectively. The relative increase in AGD using the conventional dosemeter was 11.2% in DM mode and
12.2% in DBT mode. The relative increase in AGD using NaCl pellets was 14.6% in DM mode. Our measurements suggest
that the use of simultaneous breast radiography and MI increases the AGD by 13% on average. The increase in dose is still
below the acceptable values in mammography screening recommended by the European Guidelines.
(Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/d1813ce7-6bee-49c3-9a39-4162425029f1
- author
- Bjerkén, Anna LU ; Tomic, Hanna LU ; Bernhardsson, Christian LU ; Zackrisson, Sophia LU ; Tingberg, Anders LU ; Dustler, Magnus LU and Bakic, Predrag LU
- organization
- publishing date
- 2023-04-28
- type
- Chapter in Book/Report/Conference proceeding
- publication status
- published
- subject
- keywords
- Digital breast tomosynthesis, Mechanical imaging, DBTMI, radiation dose measurement, OSL, NaCl pellets
- host publication
- Medical Imaging 2023 : Physics of Medical Imaging - Physics of Medical Imaging
- editor
- Yu, Lifeng ; Fahrig, Rebecca and M. Sabol, John
- volume
- 12463
- article number
- 124634D
- publisher
- SPIE
- conference name
- SPIE Medical Imaging 2023
- conference dates
- 2023-02-19 - 2023-02-23
- external identifiers
-
- scopus:85160752941
- DOI
- 10.1117/12.2655776
- project
- AI together with MI in mammography and breast tomosynthesis
- Simultaneous Digital Breast Tomosynthesis and Mechanical Imaging
- language
- English
- LU publication?
- yes
- additional info
- This work was supported by grants from the European Commission H2020 Marie Skłodowska Curie Actions Fellowship (IF 846540), Cancerfonden, the Swedish Breast Cancer Association, Region Skåne (Regionalt forskningsstöd), and Stiftelsen för Cancerforskning vid Onkologiska kliniken vid Universitetssjukhuset MAS.
- id
- d1813ce7-6bee-49c3-9a39-4162425029f1
- date added to LUP
- 2023-05-16 11:30:22
- date last changed
- 2024-11-16 19:53:12
@inproceedings{d1813ce7-6bee-49c3-9a39-4162425029f1, abstract = {{This study investigates the impact in terms of radiation dose when performing simultaneous digital breast tomosynthesis<br/>(DBT) and mechanical imaging (MI) – DBTMI. DBTMI has demonstrated the potential to increase specificity of cancer<br/>detection, and reduce unnecessary biopsies, as compared to digital mammography (DM) screening. The presence of the<br/>MI sensor during simultaneous image acquisition may increase the radiation dose when automatic exposure control is used.<br/>In this project, a radiation dose study was conducted on clinically available breast imaging systems with and without the<br/>MI sensor. We have investigated three approaches to analyse the dose increase in DBTMI, using (i) the estimates of average<br/>glandular dose (AGD) reported in DICOM headers of radiography images; (ii) AGD measured by a conventional<br/>dosemeter; and (iii) AGD measured by optically stimulated luminescence using NaCl pellets. The relative increase in AGD<br/>estimated from DICOM headers when using the MI sensor was on average 10.7% and 12.4%, for DM and DBT<br/>measurements, respectively. The relative increase in AGD using the conventional dosemeter was 11.2% in DM mode and<br/>12.2% in DBT mode. The relative increase in AGD using NaCl pellets was 14.6% in DM mode. Our measurements suggest<br/>that the use of simultaneous breast radiography and MI increases the AGD by 13% on average. The increase in dose is still<br/>below the acceptable values in mammography screening recommended by the European Guidelines.<br/>}}, author = {{Bjerkén, Anna and Tomic, Hanna and Bernhardsson, Christian and Zackrisson, Sophia and Tingberg, Anders and Dustler, Magnus and Bakic, Predrag}}, booktitle = {{Medical Imaging 2023 : Physics of Medical Imaging}}, editor = {{Yu, Lifeng and Fahrig, Rebecca and M. Sabol, John}}, keywords = {{Digital breast tomosynthesis; Mechanical imaging; DBTMI; radiation dose measurement; OSL; NaCl pellets}}, language = {{eng}}, month = {{04}}, publisher = {{SPIE}}, title = {{Dose evaluation of simultaneous breast radiography and mechanical imaging}}, url = {{http://dx.doi.org/10.1117/12.2655776}}, doi = {{10.1117/12.2655776}}, volume = {{12463}}, year = {{2023}}, }