Characterisation of a pre-cell hit detector to be used in single cell irradiation experiments at the Lund Nuclear Microprobe
(2008) In Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms 266(21). p.4808-4815- Abstract
- This paper describes the characterisation of an ultra-thin silicon semicondoctor Delta E detector to be used as a pre-cell ion hit detector in single ion experiments on individual, living cells. The characteristics of interest for this specific application are the hit detection efficiency, which has to be close to 100% to enable bombardment with either a single ion or a counted number of ions, the beam spreading, which should be as small as possible to maintain the targeting accuracy, and the vacuum tightness, since the detector is intended, if possible, to be used simultaneously as vacuum window. The hit detection efficiency was shown to be above 99% when detecting alpha particles or 2 MeV protons, the increase in beam size was about 1 mu... (More)
- This paper describes the characterisation of an ultra-thin silicon semicondoctor Delta E detector to be used as a pre-cell ion hit detector in single ion experiments on individual, living cells. The characteristics of interest for this specific application are the hit detection efficiency, which has to be close to 100% to enable bombardment with either a single ion or a counted number of ions, the beam spreading, which should be as small as possible to maintain the targeting accuracy, and the vacuum tightness, since the detector is intended, if possible, to be used simultaneously as vacuum window. The hit detection efficiency was shown to be above 99% when detecting alpha particles or 2 MeV protons, the increase in beam size was about 1 mu m and the vacuum tightness was comparable to that of the Si3N4 wafer which is normally used as vacuum window, thus the Delta E detector fulfils the main criteria to function properly as a single ion hit detector. (C) 2008 Elsevier B.V. All rights reserved. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1307489
- author
- Nilsson, Charlotta LU ; Pallon, Jan LU ; Thungstrom, Goran ; Arteaga, Natalia LU ; Elfman, Mikael LU ; Kristiansson, Per LU ; Nilsson, Christer LU and Wegdén, Marie LU
- organization
- publishing date
- 2008
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- Detector, Single ions, Thin silicon detector, Delta E detector, Vacuum window, efficiency
- in
- Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms
- volume
- 266
- issue
- 21
- pages
- 4808 - 4815
- publisher
- Elsevier
- external identifiers
-
- wos:000261263700013
- scopus:54049101943
- ISSN
- 0168-583X
- DOI
- 10.1016/j.nimb.2008.07.011
- language
- English
- LU publication?
- yes
- additional info
- The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Nuclear Physics (Faculty of Technology) (011013007)
- id
- 10000038-cdbc-4d3f-90ab-788d54754153 (old id 1307489)
- date added to LUP
- 2016-04-01 13:04:31
- date last changed
- 2022-01-27 17:12:03
@article{10000038-cdbc-4d3f-90ab-788d54754153, abstract = {{This paper describes the characterisation of an ultra-thin silicon semicondoctor Delta E detector to be used as a pre-cell ion hit detector in single ion experiments on individual, living cells. The characteristics of interest for this specific application are the hit detection efficiency, which has to be close to 100% to enable bombardment with either a single ion or a counted number of ions, the beam spreading, which should be as small as possible to maintain the targeting accuracy, and the vacuum tightness, since the detector is intended, if possible, to be used simultaneously as vacuum window. The hit detection efficiency was shown to be above 99% when detecting alpha particles or 2 MeV protons, the increase in beam size was about 1 mu m and the vacuum tightness was comparable to that of the Si3N4 wafer which is normally used as vacuum window, thus the Delta E detector fulfils the main criteria to function properly as a single ion hit detector. (C) 2008 Elsevier B.V. All rights reserved.}}, author = {{Nilsson, Charlotta and Pallon, Jan and Thungstrom, Goran and Arteaga, Natalia and Elfman, Mikael and Kristiansson, Per and Nilsson, Christer and Wegdén, Marie}}, issn = {{0168-583X}}, keywords = {{Detector; Single ions; Thin silicon detector; Delta E detector; Vacuum window; efficiency}}, language = {{eng}}, number = {{21}}, pages = {{4808--4815}}, publisher = {{Elsevier}}, series = {{Nuclear Instruments & Methods in Physics Research. Section B: Beam Interactions with Materials and Atoms}}, title = {{Characterisation of a pre-cell hit detector to be used in single cell irradiation experiments at the Lund Nuclear Microprobe}}, url = {{http://dx.doi.org/10.1016/j.nimb.2008.07.011}}, doi = {{10.1016/j.nimb.2008.07.011}}, volume = {{266}}, year = {{2008}}, }