Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

A new general-purpose isothermal microcalorimeter for use at temperatures up to 200 degrees C

Wadsö, Ingemar LU and Wadsö, Lars LU (2003) In Thermochimica Acta 405(1). p.15-20
Abstract
The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 degreesC. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use

f an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the... (More)
The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 degreesC. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use

f an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the temperature range 15-200 degreesC. Typical values obtained during 10 h periods at 200 degreesC are +/-3 and +/-10 nW for the baseline drift and baseline fluctuations, respectively. The heat detection limit, determined by release of electrical energy, is about 2 muJ. Preliminary stability measurements have been conducted at 100 degreesC on samples of stabilised and non-stabilised polyamide film. (C) 2003 Elsevier Science B.V. All rights reserved. (Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Isothermal microcalorimeter, Nanowatt, 200 °C, Stability measurements
in
Thermochimica Acta
volume
405
issue
1
pages
15 - 20
publisher
Elsevier
external identifiers
  • wos:000186127000003
  • scopus:0344033884
ISSN
0040-6031
DOI
10.1016/S0040-6031(03)00124-2
language
English
LU publication?
yes
id
59bcde93-ba0b-454a-b698-b8532ec3cc32 (old id 122254)
date added to LUP
2016-04-01 15:25:55
date last changed
2022-01-28 05:16:16
@article{59bcde93-ba0b-454a-b698-b8532ec3cc32,
  abstract     = {{The design and some properties of a new general-purpose isothermal microcalorimeter are reported. The instrument is a twin thermopile heat conduction calorimeter, which is designed for use up to 200 degreesC. The calorimetric units and surrounding heat sink are suspended inside a hollow aluminium construction, which is thermostated. Above that unit a second thermostated block is positioned and the whole assembly is suspended inside a Dewar vessel. When the instrument is used at room temperature and below, the thermostated units are cooled by use <br/><br>
f an insertion Peltier effect cooler. The instrument can be used with a wide range of different reaction vessels (diameter 14 mm). Baseline experiments have been conducted in the temperature range 15-200 degreesC. Typical values obtained during 10 h periods at 200 degreesC are +/-3 and +/-10 nW for the baseline drift and baseline fluctuations, respectively. The heat detection limit, determined by release of electrical energy, is about 2 muJ. Preliminary stability measurements have been conducted at 100 degreesC on samples of stabilised and non-stabilised polyamide film. (C) 2003 Elsevier Science B.V. All rights reserved.}},
  author       = {{Wadsö, Ingemar and Wadsö, Lars}},
  issn         = {{0040-6031}},
  keywords     = {{Isothermal microcalorimeter; Nanowatt; 200 °C; Stability measurements}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{15--20}},
  publisher    = {{Elsevier}},
  series       = {{Thermochimica Acta}},
  title        = {{A new general-purpose isothermal microcalorimeter for use at temperatures up to 200 degrees C}},
  url          = {{http://dx.doi.org/10.1016/S0040-6031(03)00124-2}},
  doi          = {{10.1016/S0040-6031(03)00124-2}},
  volume       = {{405}},
  year         = {{2003}},
}