Advanced

Staged proximal deployment of the Zenith TX2 thoracic stent-graft: a novel technique to improve conformance to the aortic arch.

Kölbel, Tilo LU ; Resch, Timothy A; Dias, Nuno LU ; Björses, Katarina LU ; Sonesson, Björn LU and Malina, Martin LU (2009) In Journal of Endovascular Therapy 16(5). p.598-602
Abstract
PURPOSE: To present a modification of the Zenith TX2 thoracic stent-graft that allows staged proximal deployment to improve apposition to the aortic wall. TECHNIQUE: Three standard Zenith TX2 thoracic stent-grafts and 3 modified versions were deployed in a glass model of the aortic arch. Deployment sequences were analyzed. In a patient with a 6-cm thoracic aortic aneurysm after a type B dissection, the modified Zenith TX2 thoracic stent-graft was deployed successfully and without complications; the proximal part of the stent-graft protruded less into the arch, significantly improving wall apposition. CONCLUSION: Staged proximal deployment with a modified Zenith TX2 thoracic stent-graft can improve orientation and wall apposition of the... (More)
PURPOSE: To present a modification of the Zenith TX2 thoracic stent-graft that allows staged proximal deployment to improve apposition to the aortic wall. TECHNIQUE: Three standard Zenith TX2 thoracic stent-grafts and 3 modified versions were deployed in a glass model of the aortic arch. Deployment sequences were analyzed. In a patient with a 6-cm thoracic aortic aneurysm after a type B dissection, the modified Zenith TX2 thoracic stent-graft was deployed successfully and without complications; the proximal part of the stent-graft protruded less into the arch, significantly improving wall apposition. CONCLUSION: Staged proximal deployment with a modified Zenith TX2 thoracic stent-graft can improve orientation and wall apposition of the first Z stent. A better proximal apposition may prevent early and late stent-graft complications. (Less)
Please use this url to cite or link to this publication:
author
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Journal of Endovascular Therapy
volume
16
issue
5
pages
598 - 602
publisher
International Society of Endovascular Specialists
external identifiers
  • wos:000271308800010
  • pmid:19842728
  • scopus:71949109190
ISSN
1545-1550
DOI
10.1583/09-2787.1
language
English
LU publication?
yes
id
7a6cd082-ff15-47d0-85c6-592e98819b32 (old id 1500177)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/19842728?dopt=Abstract
date added to LUP
2009-11-03 10:19:55
date last changed
2017-01-01 07:40:22
@article{7a6cd082-ff15-47d0-85c6-592e98819b32,
  abstract     = {PURPOSE: To present a modification of the Zenith TX2 thoracic stent-graft that allows staged proximal deployment to improve apposition to the aortic wall. TECHNIQUE: Three standard Zenith TX2 thoracic stent-grafts and 3 modified versions were deployed in a glass model of the aortic arch. Deployment sequences were analyzed. In a patient with a 6-cm thoracic aortic aneurysm after a type B dissection, the modified Zenith TX2 thoracic stent-graft was deployed successfully and without complications; the proximal part of the stent-graft protruded less into the arch, significantly improving wall apposition. CONCLUSION: Staged proximal deployment with a modified Zenith TX2 thoracic stent-graft can improve orientation and wall apposition of the first Z stent. A better proximal apposition may prevent early and late stent-graft complications.},
  author       = {Kölbel, Tilo and Resch, Timothy A and Dias, Nuno and Björses, Katarina and Sonesson, Björn and Malina, Martin},
  issn         = {1545-1550},
  language     = {eng},
  number       = {5},
  pages        = {598--602},
  publisher    = {International Society of Endovascular Specialists},
  series       = {Journal of Endovascular Therapy},
  title        = {Staged proximal deployment of the Zenith TX2 thoracic stent-graft: a novel technique to improve conformance to the aortic arch.},
  url          = {http://dx.doi.org/10.1583/09-2787.1},
  volume       = {16},