3D geometric gel dosimetry verification of intraprostatic fiducial guided hypofractionated radiotherapy of prostate cancer

Nordström, Fredrik LU ; Ceberg, Sofie LU ; af Wetterstedt, Sacha ; Nilsson, Per LU orcid ; Ceberg, Crister LU orcid and Bäck, Sven LU (2010) IC3DDose: The 6th International Conference on 3D Radiation Dosimetry 250.
Abstract
This pre-study is aimed to investigate the feasibility of a normoxic polyacrylamide gel (nPAG) dosimeter with implanted gold fiducials to evaluate the geometric precision, including setup correction strategies, in the delivery of hypofractionated treatments. For this purpose a phantom consisting of three parts was constructed: (1) the patient simulating volume, providing realistic scatter conditions and weight, (2) a bottle containing the active dosimetric volume and (3) the gold fiducials and the fiducial support structure. A 6.1 Gy prostate IMRT treatment was delivered to the phantom using the sliding-window technique. The phantom was positioned prior to the treatment using the implanted fiducials and kV on-board imaging. An overlay of... (More)
This pre-study is aimed to investigate the feasibility of a normoxic polyacrylamide gel (nPAG) dosimeter with implanted gold fiducials to evaluate the geometric precision, including setup correction strategies, in the delivery of hypofractionated treatments. For this purpose a phantom consisting of three parts was constructed: (1) the patient simulating volume, providing realistic scatter conditions and weight, (2) a bottle containing the active dosimetric volume and (3) the gold fiducials and the fiducial support structure. A 6.1 Gy prostate IMRT treatment was delivered to the phantom using the sliding-window technique. The phantom was positioned prior to the treatment using the implanted fiducials and kV on-board imaging. An overlay of the 95% isosurface of the TPS calculated dose distribution and the measured dose distribution using gel showed good agreement. The clinical target volume (CTV) was well centred inside the 95% isodose surface of the measured volume. It was shown for the evaluated case that the use of on-board imaging and integrated setup correction tools could be used to compensate for a deliberately introduced offset in CTV position. The study showed that MRI based nPAG gel dosimetry can be used to verify setup correction procedures using implanted gold fiducials. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
keywords
3D verification, radiotherapy, polymer gel
host publication
Journal of Physics: Conference Series
volume
250
publisher
IOP Publishing
conference name
IC3DDose: The 6th International Conference on 3D Radiation Dosimetry
conference location
Hilton Head Island, South Carolina, United States
conference dates
2010-08-22
external identifiers
  • scopus:78650535940
DOI
10.1088/1742-6596/250/1/012059
language
English
LU publication?
yes
id
cbe399c6-8873-4772-beeb-089e7fb86709 (old id 2213096)
alternative location
http://iopscience.iop.org/1742-6596/250/1/012059
date added to LUP
2016-04-04 11:59:33
date last changed
2022-01-29 22:47:10
@inproceedings{cbe399c6-8873-4772-beeb-089e7fb86709,
  abstract     = {{This pre-study is aimed to investigate the feasibility of a normoxic polyacrylamide gel (nPAG) dosimeter with implanted gold fiducials to evaluate the geometric precision, including setup correction strategies, in the delivery of hypofractionated treatments. For this purpose a phantom consisting of three parts was constructed: (1) the patient simulating volume, providing realistic scatter conditions and weight, (2) a bottle containing the active dosimetric volume and (3) the gold fiducials and the fiducial support structure. A 6.1 Gy prostate IMRT treatment was delivered to the phantom using the sliding-window technique. The phantom was positioned prior to the treatment using the implanted fiducials and kV on-board imaging. An overlay of the 95% isosurface of the TPS calculated dose distribution and the measured dose distribution using gel showed good agreement. The clinical target volume (CTV) was well centred inside the 95% isodose surface of the measured volume. It was shown for the evaluated case that the use of on-board imaging and integrated setup correction tools could be used to compensate for a deliberately introduced offset in CTV position. The study showed that MRI based nPAG gel dosimetry can be used to verify setup correction procedures using implanted gold fiducials.}},
  author       = {{Nordström, Fredrik and Ceberg, Sofie and af Wetterstedt, Sacha and Nilsson, Per and Ceberg, Crister and Bäck, Sven}},
  booktitle    = {{Journal of Physics: Conference Series}},
  keywords     = {{3D verification; radiotherapy; polymer gel}},
  language     = {{eng}},
  publisher    = {{IOP Publishing}},
  title        = {{3D geometric gel dosimetry verification of intraprostatic fiducial guided hypofractionated radiotherapy of prostate cancer}},
  url          = {{http://dx.doi.org/10.1088/1742-6596/250/1/012059}},
  doi          = {{10.1088/1742-6596/250/1/012059}},
  volume       = {{250}},
  year         = {{2010}},
}