Theoretical analysis of transurethral laser-induced thermo-therapy for treatment of benign prostatic hyperplasia. Evaluation of a water-cooled applicator

Sturesson, C and Andersson-Engels, Stefan LU (1996) In Physics in Medicine and Biology 41(3). p.445-463
Abstract
A mathematical model for predicting the temperature rise in transurethral laser-induced thermo-therapy for benign prostatic hyperplasia was developed. In the model an optical line source emitting light from an Nd:YAG laser isotropically was placed in the urethra. Water cooling of the urethral epithelium was modelled using a two-tube system. The relationship between the difference in outlet and inlet water temperatures and the highest tissue temperature level reached was theoretically investigated. It was found that the water temperature difference was linearly dependent on the steady-state maximum tissue temperature. The theoretical calculations suggest that the water-cooled applicator can be used to measure the maximum tissue temperature.... (More)
A mathematical model for predicting the temperature rise in transurethral laser-induced thermo-therapy for benign prostatic hyperplasia was developed. In the model an optical line source emitting light from an Nd:YAG laser isotropically was placed in the urethra. Water cooling of the urethral epithelium was modelled using a two-tube system. The relationship between the difference in outlet and inlet water temperatures and the highest tissue temperature level reached was theoretically investigated. It was found that the water temperature difference was linearly dependent on the steady-state maximum tissue temperature. The theoretical calculations suggest that the water-cooled applicator can be used to measure the maximum tissue temperature. With temperature control, the prostatic tissue temperature can be prevented from exceeding the boiling point of water, excluding tissue carbonization. The model was also used to evaluate the influence of a number of different parameters on the damaged tissue volume. Increasing the urethral lumen radius by a factor of two by means of inserting different sized tubes was found to augment the tissue volume raised to therapeutic temperatures by up to 50%. The calculations showed that cooling of the urethral epithelium can result in an increase in the damaged volume by 80% as compared to not applying any cooling. The temperature of the cooling water was found to influence the tissue temperature only to a small extent. (Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Physics in Medicine and Biology
volume
41
issue
3
pages
445 - 463
publisher
IOP Publishing
external identifiers
  • scopus:0029920381
ISSN
1361-6560
DOI
10.1088/0031-9155/41/3/008
language
English
LU publication?
yes
id
04b2a27b-e872-4559-adae-b9c7399d13f2 (old id 2259432)
date added to LUP
2016-04-04 08:08:25
date last changed
2022-03-07 21:13:12
@article{04b2a27b-e872-4559-adae-b9c7399d13f2,
  abstract     = {{A mathematical model for predicting the temperature rise in transurethral laser-induced thermo-therapy for benign prostatic hyperplasia was developed. In the model an optical line source emitting light from an Nd:YAG laser isotropically was placed in the urethra. Water cooling of the urethral epithelium was modelled using a two-tube system. The relationship between the difference in outlet and inlet water temperatures and the highest tissue temperature level reached was theoretically investigated. It was found that the water temperature difference was linearly dependent on the steady-state maximum tissue temperature. The theoretical calculations suggest that the water-cooled applicator can be used to measure the maximum tissue temperature. With temperature control, the prostatic tissue temperature can be prevented from exceeding the boiling point of water, excluding tissue carbonization. The model was also used to evaluate the influence of a number of different parameters on the damaged tissue volume. Increasing the urethral lumen radius by a factor of two by means of inserting different sized tubes was found to augment the tissue volume raised to therapeutic temperatures by up to 50%. The calculations showed that cooling of the urethral epithelium can result in an increase in the damaged volume by 80% as compared to not applying any cooling. The temperature of the cooling water was found to influence the tissue temperature only to a small extent.}},
  author       = {{Sturesson, C and Andersson-Engels, Stefan}},
  issn         = {{1361-6560}},
  language     = {{eng}},
  number       = {{3}},
  pages        = {{445--463}},
  publisher    = {{IOP Publishing}},
  series       = {{Physics in Medicine and Biology}},
  title        = {{Theoretical analysis of transurethral laser-induced thermo-therapy for treatment of benign prostatic hyperplasia. Evaluation of a water-cooled applicator}},
  url          = {{https://lup.lub.lu.se/search/files/5167367/2297598.pdf}},
  doi          = {{10.1088/0031-9155/41/3/008}},
  volume       = {{41}},
  year         = {{1996}},
}