Abstract
BACKGROUND AND PURPOSE: Ensuring data integrity in radiotherapy is of major importance and a complex task. The aim of this study was to compare three different combinations of treatment planning and record and verify systems with respect to data integrity.



MATERIALS AND METHODS: A software for comparison of treatment parameters in DICOM-RT files was developed using the MATLAB R2010a (MathWorks Inc.) environment. One hundred treatment plans were analyzed for each system combination. In the first step of the analysis, all parameters were compared and a normal condition for each system combination was identified. The second step focused on the discovery of potential special cause deviations, e.g. by applying tolerance... (More)
BACKGROUND AND PURPOSE: Ensuring data integrity in radiotherapy is of major importance and a complex task. The aim of this study was to compare three different combinations of treatment planning and record and verify systems with respect to data integrity.



MATERIALS AND METHODS: A software for comparison of treatment parameters in DICOM-RT files was developed using the MATLAB R2010a (MathWorks Inc.) environment. One hundred treatment plans were analyzed for each system combination. In the first step of the analysis, all parameters were compared and a normal condition for each system combination was identified. The second step focused on the discovery of potential special cause deviations, e.g. by applying tolerance levels.



RESULTS: In total, 15% and 0.37% of all comparisons failed to meet the defined integrity demands in step 1 and step 2 of the analysis, respectively. Differences in the data integrity level between the systems were observed, ranging on average from 3.1 to 11.9 discrepancies per beam for the different RV-TPS combinations.



CONCLUSIONS: The proposed method can be used to increase the safety for individual patients by ensuring that the intended treatment is delivered. The system combination with the highest level of data integrity was found to be the one which shares a single database. (Less)
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Radiotherapy and Oncology
volume
103
issue
3
pages
299 - 304
publisher
Elsevier
external identifiers
  • wos:000305845000004
  • pmid:22342418
  • scopus:84861823622
  • pmid:22342418
ISSN
1879-0887
DOI
10.1016/j.radonc.2012.01.004
language
English
LU publication?
yes
id
47a48d6a-e1b5-4613-b763-9d15e1e173d7 (old id 2366788)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/22342418?dopt=Abstract
date added to LUP
2016-04-01 09:56:59
date last changed
2022-01-25 18:15:28
@article{47a48d6a-e1b5-4613-b763-9d15e1e173d7,
  abstract     = {{BACKGROUND AND PURPOSE: Ensuring data integrity in radiotherapy is of major importance and a complex task. The aim of this study was to compare three different combinations of treatment planning and record and verify systems with respect to data integrity. <br/><br>
<br/><br>
MATERIALS AND METHODS: A software for comparison of treatment parameters in DICOM-RT files was developed using the MATLAB R2010a (MathWorks Inc.) environment. One hundred treatment plans were analyzed for each system combination. In the first step of the analysis, all parameters were compared and a normal condition for each system combination was identified. The second step focused on the discovery of potential special cause deviations, e.g. by applying tolerance levels. <br/><br>
<br/><br>
RESULTS: In total, 15% and 0.37% of all comparisons failed to meet the defined integrity demands in step 1 and step 2 of the analysis, respectively. Differences in the data integrity level between the systems were observed, ranging on average from 3.1 to 11.9 discrepancies per beam for the different RV-TPS combinations. <br/><br>
<br/><br>
CONCLUSIONS: The proposed method can be used to increase the safety for individual patients by ensuring that the intended treatment is delivered. The system combination with the highest level of data integrity was found to be the one which shares a single database.}},
  author       = {{Nordström, Fredrik and Ceberg, Crister and Bäck, Sven}},
  issn         = {{1879-0887}},
  language     = {{eng}},
  number       = {{3}},
  pages        = {{299--304}},
  publisher    = {{Elsevier}},
  series       = {{Radiotherapy and Oncology}},
  title        = {{Ensuring the integrity of treatment parameters throughout the radiotherapy process.}},
  url          = {{http://dx.doi.org/10.1016/j.radonc.2012.01.004}},
  doi          = {{10.1016/j.radonc.2012.01.004}},
  volume       = {{103}},
  year         = {{2012}},
}