Dual-window scatter correction and energy window setting in cerebral blood flow SPECT: a Monte Carlo study

Gustafsson, A ; Arlig, A ; Jacobsson, L ; Ljungberg, Michael LU and Wikkelso, C (2000) In Physics in Medicine and Biology 45(11). p.3431-3440
Abstract
The image quality in SPECT studies of the regional cerebral blood flow (rCBF) performed with 99mTc-HMPAO is degraded by scattered photons. The finite energy resolution of the gamma camera makes the detection of scattered photons unavoidable, and this is observed in the image as an impaired contrast between grey and white matter structures. In this work, a Monte Carlo simulated SPECT study of a realistic voxel-based brain phantom was used to evaluate the resulting contrast-to-noise ratio for a number of energy window settings, with and without the dual-window scatter correction. Values of the scaling factor k, used to obtain the fraction of scattered photons in the photopeak window, were estimated for each energy window. The use of a... (More)
The image quality in SPECT studies of the regional cerebral blood flow (rCBF) performed with 99mTc-HMPAO is degraded by scattered photons. The finite energy resolution of the gamma camera makes the detection of scattered photons unavoidable, and this is observed in the image as an impaired contrast between grey and white matter structures. In this work, a Monte Carlo simulated SPECT study of a realistic voxel-based brain phantom was used to evaluate the resulting contrast-to-noise ratio for a number of energy window settings, with and without the dual-window scatter correction. Values of the scaling factor k, used to obtain the fraction of scattered photons in the photopeak window, were estimated for each energy window. The use of a narrower, asymmetric, energy discrimination window improved the contrast, with a subsequent increase in statistical noise due to the lower number of counts. The photopeak-window setting giving the best contrast-to-noise ratio was found to be the same whether or not scatter correction was applied. Its value was 17% centred at 142 keV. At the optimum photopeak-window setting, the contrast was improved by using scatter correction, but the contrast-to-noise ratio was made worse. (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Physics in Medicine and Biology
volume
45
issue
11
pages
3431 - 3440
publisher
IOP Publishing
external identifiers
  • pmid:11098915
  • scopus:0033769669
ISSN
1361-6560
DOI
10.1088/0031-9155/45/11/323
language
English
LU publication?
yes
id
42ab1f4e-82d3-49ca-8de6-77d38c34e804 (old id 1116458)
date added to LUP
2016-04-01 12:06:29
date last changed
2022-01-26 22:51:57
@article{42ab1f4e-82d3-49ca-8de6-77d38c34e804,
  abstract     = {{The image quality in SPECT studies of the regional cerebral blood flow (rCBF) performed with 99mTc-HMPAO is degraded by scattered photons. The finite energy resolution of the gamma camera makes the detection of scattered photons unavoidable, and this is observed in the image as an impaired contrast between grey and white matter structures. In this work, a Monte Carlo simulated SPECT study of a realistic voxel-based brain phantom was used to evaluate the resulting contrast-to-noise ratio for a number of energy window settings, with and without the dual-window scatter correction. Values of the scaling factor k, used to obtain the fraction of scattered photons in the photopeak window, were estimated for each energy window. The use of a narrower, asymmetric, energy discrimination window improved the contrast, with a subsequent increase in statistical noise due to the lower number of counts. The photopeak-window setting giving the best contrast-to-noise ratio was found to be the same whether or not scatter correction was applied. Its value was 17% centred at 142 keV. At the optimum photopeak-window setting, the contrast was improved by using scatter correction, but the contrast-to-noise ratio was made worse.}},
  author       = {{Gustafsson, A and Arlig, A and Jacobsson, L and Ljungberg, Michael and Wikkelso, C}},
  issn         = {{1361-6560}},
  language     = {{eng}},
  number       = {{11}},
  pages        = {{3431--3440}},
  publisher    = {{IOP Publishing}},
  series       = {{Physics in Medicine and Biology}},
  title        = {{Dual-window scatter correction and energy window setting in cerebral blood flow SPECT: a Monte Carlo study}},
  url          = {{http://dx.doi.org/10.1088/0031-9155/45/11/323}},
  doi          = {{10.1088/0031-9155/45/11/323}},
  volume       = {{45}},
  year         = {{2000}},
}