Development and Implementation of Cardiac Event Detectors in Digital CMOS
(2005)- Abstract
- This doctoral dissertation presents the development and digital hardware realization of cardiac event detectors. Implantable medical
appliances, as the cardiac pacemaker, have progressed from a life
sustaining device to a device that considerably improves life
quality for all ages. The number of electronic devices and household
appliances in everyday live has an ongoing exponential growth. These
devices contaminate their environment with electronic, magnetic or
electromagnetic radiation. Pacemaker patients exposed to this
environment may suffer due to malfunction of the pacemaker. Thus,
the... (More) - This doctoral dissertation presents the development and digital hardware realization of cardiac event detectors. Implantable medical
appliances, as the cardiac pacemaker, have progressed from a life
sustaining device to a device that considerably improves life
quality for all ages. The number of electronic devices and household
appliances in everyday live has an ongoing exponential growth. These
devices contaminate their environment with electronic, magnetic or
electromagnetic radiation. Pacemaker patients exposed to this
environment may suffer due to malfunction of the pacemaker. Thus,
the next generation of pacemakers require a low-power consuming
event detector that provides reliable detection performance.
In this thesis two papers that present an artificial neural network
based event detector for R-wave detection are merged to an extended
manuscript. The neural network functions as a whitening filter prior
to a matched filter. It is shown how the neural network responds to
sudden changes in the input sequence. An algorithm that determines
the initial template for matched filtering is proposed, and a
continuous update of the filter impulse response is implemented in
order to track long-term changes in signal morphology. Furthermore,
an updated threshold function is proposed which addresses amplitude
variations in the electrogram. Noise suppression and classification
performance under ``real-life situation'' are explored by analyzing
recordings from databases of electrograms and noise. Finally, the
suitability for pacemaker application is discussed.
Four papers that present a low-power digital hardware implementation
of a wavelet based event detector are merged and extended in the
second part of this thesis. The theory of the wavelet filterbank is
presented, and it is shown how the architecture was modified to
achieve an area and power efficient silicon implementation. An
algorithm is presented that determines automatically a threshold
level during the initialization phase. A second operation mode is
proposed to shut down major parts of the hardware, if the patient is
at rest or in a ``low-noise'' environment. Power analysis on
RTL-level shows that leakage power is the dominant factor in the
total power figure. An estimate for leakage reduction is presented
if sleep transistors are introduced between the supply rails and the
logic that is shut-off in low-noise operation mode. The R-wave
detector has been implemented in 0.13$,mu$m low-leakage CMOS
technology. The design has been routed, and, thereafter, sleep
transistors are introduced in the layout. Detection performance is
evaluated by means of databases containing electrograms to which
five types of exogenic and endogenic interference are added. The
results show that reliable detection is obtained at moderate and low
SNRs. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/545396
- author
- Rodrigues, Joachim LU
- supervisor
-
- Viktor Öwall LU
- opponent
-
- Professor Lande, Tor Sverre, Dept. of Informatics, University of Oslo
- organization
- publishing date
- 2005
- type
- Thesis
- publication status
- published
- subject
- keywords
- Signal processing, Elektronik och elektroteknik, Electronics and Electrical technology, event detection, cardiac pacemaker, sleep- transistors, low-power, ASIC, Neural Networks, Electrogram, Signalbehandling, Wavelets
- pages
- 174 pages
- publisher
- Elektrovetenskap
- defense location
- Room E:1406, E-building, Ole Römers väg 3, Lund Institute of Technology
- defense date
- 2005-10-07 10:15:00
- ISBN
- 1402866256
- language
- Swedish
- LU publication?
- yes
- id
- da485c66-43f0-48c8-96fe-b0bcce24573b (old id 545396)
- date added to LUP
- 2016-04-04 10:00:54
- date last changed
- 2018-11-21 20:56:13
@phdthesis{da485c66-43f0-48c8-96fe-b0bcce24573b, abstract = {{This doctoral dissertation presents the development and digital hardware realization of cardiac event detectors. Implantable medical<br/><br> <br/><br> appliances, as the cardiac pacemaker, have progressed from a life<br/><br> <br/><br> sustaining device to a device that considerably improves life<br/><br> <br/><br> quality for all ages. The number of electronic devices and household<br/><br> <br/><br> appliances in everyday live has an ongoing exponential growth. These<br/><br> <br/><br> devices contaminate their environment with electronic, magnetic or<br/><br> <br/><br> electromagnetic radiation. Pacemaker patients exposed to this<br/><br> <br/><br> environment may suffer due to malfunction of the pacemaker. Thus,<br/><br> <br/><br> the next generation of pacemakers require a low-power consuming<br/><br> <br/><br> event detector that provides reliable detection performance.<br/><br> <br/><br> In this thesis two papers that present an artificial neural network<br/><br> <br/><br> based event detector for R-wave detection are merged to an extended<br/><br> <br/><br> manuscript. The neural network functions as a whitening filter prior<br/><br> <br/><br> to a matched filter. It is shown how the neural network responds to<br/><br> <br/><br> sudden changes in the input sequence. An algorithm that determines<br/><br> <br/><br> the initial template for matched filtering is proposed, and a<br/><br> <br/><br> continuous update of the filter impulse response is implemented in<br/><br> <br/><br> order to track long-term changes in signal morphology. Furthermore,<br/><br> <br/><br> an updated threshold function is proposed which addresses amplitude<br/><br> <br/><br> variations in the electrogram. Noise suppression and classification<br/><br> <br/><br> performance under ``real-life situation'' are explored by analyzing<br/><br> <br/><br> recordings from databases of electrograms and noise. Finally, the<br/><br> <br/><br> suitability for pacemaker application is discussed.<br/><br> <br/><br> Four papers that present a low-power digital hardware implementation<br/><br> <br/><br> of a wavelet based event detector are merged and extended in the<br/><br> <br/><br> second part of this thesis. The theory of the wavelet filterbank is<br/><br> <br/><br> presented, and it is shown how the architecture was modified to<br/><br> <br/><br> achieve an area and power efficient silicon implementation. An<br/><br> <br/><br> algorithm is presented that determines automatically a threshold<br/><br> <br/><br> level during the initialization phase. A second operation mode is<br/><br> <br/><br> proposed to shut down major parts of the hardware, if the patient is<br/><br> <br/><br> at rest or in a ``low-noise'' environment. Power analysis on<br/><br> <br/><br> RTL-level shows that leakage power is the dominant factor in the<br/><br> <br/><br> total power figure. An estimate for leakage reduction is presented<br/><br> <br/><br> if sleep transistors are introduced between the supply rails and the<br/><br> <br/><br> logic that is shut-off in low-noise operation mode. The R-wave<br/><br> <br/><br> detector has been implemented in 0.13$,mu$m low-leakage CMOS<br/><br> <br/><br> technology. The design has been routed, and, thereafter, sleep<br/><br> <br/><br> transistors are introduced in the layout. Detection performance is<br/><br> <br/><br> evaluated by means of databases containing electrograms to which<br/><br> <br/><br> five types of exogenic and endogenic interference are added. The<br/><br> <br/><br> results show that reliable detection is obtained at moderate and low<br/><br> <br/><br> SNRs.}}, author = {{Rodrigues, Joachim}}, isbn = {{1402866256}}, keywords = {{Signal processing; Elektronik och elektroteknik; Electronics and Electrical technology; event detection; cardiac pacemaker; sleep- transistors; low-power; ASIC; Neural Networks; Electrogram; Signalbehandling; Wavelets}}, language = {{swe}}, publisher = {{Elektrovetenskap}}, school = {{Lund University}}, title = {{Development and Implementation of Cardiac Event Detectors in Digital CMOS}}, year = {{2005}}, }