Attefal Dosin® -a new pump device. Its construction and function
(2007) In Uppsala Journal of Medical Sciences 112(1). p.83-93- Abstract
- The cyclic flow of the Dosin((R)) is useful in interventional radiology since it prevents clotting and increases the mixing with the blood. The flow rate varies from 0.03 mL/s to 5 mL/s and it is recommended to use fluid with a viscosity lower than 3 mPa.s. The outlet pressure is recommended below 300 mm Hg. The Dosin((R)) must be replaced after each patient, alternatively after maximum 24h continuous use.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/572083
- author
- Wahlberg, A. ; Erikson, U. and Fuchs, Laszlo LU
- organization
- publishing date
- 2007
- type
- Contribution to journal
- publication status
- published
- subject
- in
- Uppsala Journal of Medical Sciences
- volume
- 112
- issue
- 1
- pages
- 83 - 93
- publisher
- Upsala Medical Society
- external identifiers
-
- wos:000253451900008
- scopus:48949089194
- ISSN
- 0300-9734
- language
- English
- LU publication?
- yes
- id
- 94f7cec5-4a2d-4c44-bdcf-20881a934625 (old id 572083)
- date added to LUP
- 2016-04-01 11:34:15
- date last changed
- 2025-01-14 10:40:40
@article{94f7cec5-4a2d-4c44-bdcf-20881a934625, abstract = {{The cyclic flow of the Dosin((R)) is useful in interventional radiology since it prevents clotting and increases the mixing with the blood. The flow rate varies from 0.03 mL/s to 5 mL/s and it is recommended to use fluid with a viscosity lower than 3 mPa.s. The outlet pressure is recommended below 300 mm Hg. The Dosin((R)) must be replaced after each patient, alternatively after maximum 24h continuous use.}}, author = {{Wahlberg, A. and Erikson, U. and Fuchs, Laszlo}}, issn = {{0300-9734}}, language = {{eng}}, number = {{1}}, pages = {{83--93}}, publisher = {{Upsala Medical Society}}, series = {{Uppsala Journal of Medical Sciences}}, title = {{Attefal Dosin® -a new pump device. Its construction and function}}, volume = {{112}}, year = {{2007}}, }