Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

Instrumented platform for assessment of isometric hand muscles contractions

Malešević, Nebojša LU ; Andersson, Gert LU ; Björkman, Anders LU ; Controzzi, Marco ; Cipriani, Christian and Antfolk, Christian LU (2019) In Measurement Science and Technology 30(6).
Abstract

Measurement of forces exerted by a human hand while performing common gestures is a highly valuable task for assessment of neurorehabilitation and neurological disorders, but also, for control of movement that could be directly transferred to assistive devices. Even though accurate and selective multi-joint measurement of hand forces is desirable in both clinical and research applications there is no commercially available device able to perform such measurements. Moreover, the custom-made systems used in research commonly impose limitations, such as availability of only single, predefined hand aperture. Furthermore, there is no consensus on design requirements for custom made measurement systems that would enable comparison of results... (More)

Measurement of forces exerted by a human hand while performing common gestures is a highly valuable task for assessment of neurorehabilitation and neurological disorders, but also, for control of movement that could be directly transferred to assistive devices. Even though accurate and selective multi-joint measurement of hand forces is desirable in both clinical and research applications there is no commercially available device able to perform such measurements. Moreover, the custom-made systems used in research commonly impose limitations, such as availability of only single, predefined hand aperture. Furthermore, there is no consensus on design requirements for custom made measurement systems that would enable comparison of results obtained during research or clinical hand function studies. In an attempt to provide a possible solution for a device capable of multi-joint hand forces measurement and disseminate it to the research community, this paper presents the mechanical and electronic design of an instrumented platform for assessment of isometric hand muscles contractions. Some of the key features related to the developed system are: flexibility in placing the hand/fingers, fast and easy hand fitting, adjustability to different lengths, circumferences and postures of the digits, and the possibility to register individual bidirectional forces from the digits and the wrist. The accuracy of isometric force measurements was evaluated in a controlled test with the reference high accuracy force gauge device during which the developed system showed high linearity (R 2 = 0.9999). As the more realistic test, the device was evaluated when force was applied to individual sensors but also during the intramuscular electromyography (iEMG) study. The data gathered during the iEMG measurements was thoroughly assessed to obtain three appropriate metrics; the first estimating crosstalk between individual force sensors; the second evaluating agreement between measured forces and forces estimated through iEMG; and the third providing qualitative evaluation of hand force in respect to activations of individual muscle units. The results of these analyses performed on multiple joint forces show agreement with previously published results, but with the difference that in that case, the measurement was performed with a single degree of freedom device.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
electromyography, hand control, hand forces, isometric force measurement
in
Measurement Science and Technology
volume
30
issue
6
article number
065701
publisher
IOP Publishing
external identifiers
  • scopus:85067580941
ISSN
0957-0233
DOI
10.1088/1361-6501/ab0eae
language
English
LU publication?
yes
id
5cafd321-d9ec-4005-8ea8-87fab244efff
date added to LUP
2019-07-05 15:36:20
date last changed
2022-10-18 16:10:49
@article{5cafd321-d9ec-4005-8ea8-87fab244efff,
  abstract     = {{<p>Measurement of forces exerted by a human hand while performing common gestures is a highly valuable task for assessment of neurorehabilitation and neurological disorders, but also, for control of movement that could be directly transferred to assistive devices. Even though accurate and selective multi-joint measurement of hand forces is desirable in both clinical and research applications there is no commercially available device able to perform such measurements. Moreover, the custom-made systems used in research commonly impose limitations, such as availability of only single, predefined hand aperture. Furthermore, there is no consensus on design requirements for custom made measurement systems that would enable comparison of results obtained during research or clinical hand function studies. In an attempt to provide a possible solution for a device capable of multi-joint hand forces measurement and disseminate it to the research community, this paper presents the mechanical and electronic design of an instrumented platform for assessment of isometric hand muscles contractions. Some of the key features related to the developed system are: flexibility in placing the hand/fingers, fast and easy hand fitting, adjustability to different lengths, circumferences and postures of the digits, and the possibility to register individual bidirectional forces from the digits and the wrist. The accuracy of isometric force measurements was evaluated in a controlled test with the reference high accuracy force gauge device during which the developed system showed high linearity (R <sup>2</sup> = 0.9999). As the more realistic test, the device was evaluated when force was applied to individual sensors but also during the intramuscular electromyography (iEMG) study. The data gathered during the iEMG measurements was thoroughly assessed to obtain three appropriate metrics; the first estimating crosstalk between individual force sensors; the second evaluating agreement between measured forces and forces estimated through iEMG; and the third providing qualitative evaluation of hand force in respect to activations of individual muscle units. The results of these analyses performed on multiple joint forces show agreement with previously published results, but with the difference that in that case, the measurement was performed with a single degree of freedom device.</p>}},
  author       = {{Malešević, Nebojša and Andersson, Gert and Björkman, Anders and Controzzi, Marco and Cipriani, Christian and Antfolk, Christian}},
  issn         = {{0957-0233}},
  keywords     = {{electromyography; hand control; hand forces; isometric force measurement}},
  language     = {{eng}},
  month        = {{04}},
  number       = {{6}},
  publisher    = {{IOP Publishing}},
  series       = {{Measurement Science and Technology}},
  title        = {{Instrumented platform for assessment of isometric hand muscles contractions}},
  url          = {{http://dx.doi.org/10.1088/1361-6501/ab0eae}},
  doi          = {{10.1088/1361-6501/ab0eae}},
  volume       = {{30}},
  year         = {{2019}},
}