Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

Concurrent measurement of perfusion parameters related to small blood vessels and cerebrospinal fluid circulation in the human brain using dynamic dual-spin-echo perfusion MRI

Cao, Di ; Sun, Yuanqi ; Li, Yinghao ; Su, Pan ; Pillai, Jay J ; Qiao, Ye ; Lu, Hanzhang ; Van Zijl, Peter C M ; Knutsson, Linda LU orcid and Hua, Jun (2023) In NMR in Biomedicine 36(10).
Abstract

Accumulating evidence from recent studies has indicated the importance of studying the interaction between the microvascular and lymphatic systems in the brain. To date, most imaging methods can only measure blood or lymphatic vessels separately, such as dynamic susceptibility contrast (DSC) MRI for blood vessels and DSC MRI-in-the-cerebrospinal fluid (CSF) (cDSC MRI) for lymphatic vessels. An approach that can measure both blood and lymphatic vessels in a single scan offers advantages such as a halved scan time and contrast dosage. This study attempts to develop one such approach by optimizing a dual-echo turbo-spin-echo sequence, termed "dynamic dual-spin-echo perfusion (DDSEP) MRI". Bloch simulations were performed to optimize the... (More)

Accumulating evidence from recent studies has indicated the importance of studying the interaction between the microvascular and lymphatic systems in the brain. To date, most imaging methods can only measure blood or lymphatic vessels separately, such as dynamic susceptibility contrast (DSC) MRI for blood vessels and DSC MRI-in-the-cerebrospinal fluid (CSF) (cDSC MRI) for lymphatic vessels. An approach that can measure both blood and lymphatic vessels in a single scan offers advantages such as a halved scan time and contrast dosage. This study attempts to develop one such approach by optimizing a dual-echo turbo-spin-echo sequence, termed "dynamic dual-spin-echo perfusion (DDSEP) MRI". Bloch simulations were performed to optimize the dual-echo sequence for the measurement of gadolinium (Gd)-induced blood and CSF signal changes using a short and a long echo time, respectively. The proposed method furnishes a T1-dominant contrast in CSF and a T2-dominant contrast in blood. MRI experiments were performed in healthy subjects to evaluate the dual-echo approach by comparing it with existing separate methods. Based on simulations, the short and long echo time were chosen around the time when blood signals show maximum difference between post- and pre-Gd scans, and the time when blood signals are completely suppressed, respectively. The proposed method showed consistent results in human brains as previous studies using separate methods. Signal changes from small blood vessels occurred faster than from lymphatic vessels after intravenous Gd injection. In conclusion, Gd-induced signal changes in blood and CSF can be detected simultaneously in healthy subjects with the proposed sequence. The temporal difference in Gd-induced signal changes from small blood and lymphatic vessels after intravenous Gd injection was confirmed using the proposed approach in the same human subjects. Results from this proof-of-concept study will be used to further optimize DDSEP MRI in subsequent studies.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
NMR in Biomedicine
volume
36
issue
10
publisher
John Wiley & Sons Inc.
external identifiers
  • scopus:85163135558
  • pmid:37308297
ISSN
0952-3480
DOI
10.1002/nbm.4984
language
English
LU publication?
yes
id
63da21ef-4e7d-4d59-b32e-e7be93112fb5
date added to LUP
2023-06-19 18:23:38
date last changed
2024-04-19 22:58:23
@article{63da21ef-4e7d-4d59-b32e-e7be93112fb5,
  abstract     = {{<p>Accumulating evidence from recent studies has indicated the importance of studying the interaction between the microvascular and lymphatic systems in the brain. To date, most imaging methods can only measure blood or lymphatic vessels separately, such as dynamic susceptibility contrast (DSC) MRI for blood vessels and DSC MRI-in-the-cerebrospinal fluid (CSF) (cDSC MRI) for lymphatic vessels. An approach that can measure both blood and lymphatic vessels in a single scan offers advantages such as a halved scan time and contrast dosage. This study attempts to develop one such approach by optimizing a dual-echo turbo-spin-echo sequence, termed "dynamic dual-spin-echo perfusion (DDSEP) MRI". Bloch simulations were performed to optimize the dual-echo sequence for the measurement of gadolinium (Gd)-induced blood and CSF signal changes using a short and a long echo time, respectively. The proposed method furnishes a T1-dominant contrast in CSF and a T2-dominant contrast in blood. MRI experiments were performed in healthy subjects to evaluate the dual-echo approach by comparing it with existing separate methods. Based on simulations, the short and long echo time were chosen around the time when blood signals show maximum difference between post- and pre-Gd scans, and the time when blood signals are completely suppressed, respectively. The proposed method showed consistent results in human brains as previous studies using separate methods. Signal changes from small blood vessels occurred faster than from lymphatic vessels after intravenous Gd injection. In conclusion, Gd-induced signal changes in blood and CSF can be detected simultaneously in healthy subjects with the proposed sequence. The temporal difference in Gd-induced signal changes from small blood and lymphatic vessels after intravenous Gd injection was confirmed using the proposed approach in the same human subjects. Results from this proof-of-concept study will be used to further optimize DDSEP MRI in subsequent studies.</p>}},
  author       = {{Cao, Di and Sun, Yuanqi and Li, Yinghao and Su, Pan and Pillai, Jay J and Qiao, Ye and Lu, Hanzhang and Van Zijl, Peter C M and Knutsson, Linda and Hua, Jun}},
  issn         = {{0952-3480}},
  language     = {{eng}},
  month        = {{06}},
  number       = {{10}},
  publisher    = {{John Wiley & Sons Inc.}},
  series       = {{NMR in Biomedicine}},
  title        = {{Concurrent measurement of perfusion parameters related to small blood vessels and cerebrospinal fluid circulation in the human brain using dynamic dual-spin-echo perfusion MRI}},
  url          = {{http://dx.doi.org/10.1002/nbm.4984}},
  doi          = {{10.1002/nbm.4984}},
  volume       = {{36}},
  year         = {{2023}},
}