Accuracy of four-dimensional phase-contrast velocity mapping for blood flow visualizations: a phantom study.

Nilsson, Anders LU ; Markenroth Bloch, Karin LU orcid ; Töger, Johannes LU ; Heiberg, Einar LU and Ståhlberg, Freddy LU (2013) In Acta Radiologica 54(6). p.663-671
Abstract
BackgroundTime-resolved three-dimensional, three-directional phase-contrast magnetic resonance velocity mapping (4D PC-MRI) is a powerful technique to depict dynamic blood flow patterns in the human body. However, the impact of phase background effects on flow visualizations has not been thoroughly studied previously, and it has not yet been experimentally demonstrated to what degree phase offsets affect flow visualizations and create errors such as inaccurate particle traces.PurposeTo quantify background phase offsets and their subsequent impact on particle trace visualizations in a 4D PC-MRI sequence. Additionally, we sought to investigate to what degree visualization errors are reduced by background phase correction.Material and... (More)
BackgroundTime-resolved three-dimensional, three-directional phase-contrast magnetic resonance velocity mapping (4D PC-MRI) is a powerful technique to depict dynamic blood flow patterns in the human body. However, the impact of phase background effects on flow visualizations has not been thoroughly studied previously, and it has not yet been experimentally demonstrated to what degree phase offsets affect flow visualizations and create errors such as inaccurate particle traces.PurposeTo quantify background phase offsets and their subsequent impact on particle trace visualizations in a 4D PC-MRI sequence. Additionally, we sought to investigate to what degree visualization errors are reduced by background phase correction.Material and MethodsA rotating phantom with a known velocity field was used to quantify background phase of 4D PC-MRI sequences accelerated with SENSE as well as different k-t BLAST speed-up factors. The deviation in end positions between particle traces in the measured velocity fields were compared before and after the application of two different phase correction methods.ResultsPhantom measurements revealed background velocity offsets up to 7 cm/s (7% of velocity encoding sensitivity) in the central slice, increasing with distance from the center. Background offsets remained constant with increasing k-t BLAST speed-up factors. End deviations of up to 5.3 mm (1.8 voxels) in the direction perpendicular to the rotating disc were found between particle traces and the seeding plane of the traces. Phase correction by subtraction of the data from the stationary phantom reduced the average deviation by up to 56%, while correcting the data-set with a first-order polynomial fit to stationary regions decreased average deviation up to 78%.ConclusionPathline visualizations can be significantly affected by background phase errors, highlighting the importance of dedicated and robust phase correction methods. Our results show that pathline deviation can be substantial if adequate phase background errors are not minimized. (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Acta Radiologica
volume
54
issue
6
pages
663 - 671
publisher
SAGE Publications
external identifiers
  • wos:000323423200014
  • pmid:23486562
  • scopus:84886617616
  • pmid:23486562
ISSN
1600-0455
DOI
10.1177/0284185113478005
language
English
LU publication?
yes
id
77e282d8-1f81-4a5a-afe0-8210a7cbb999 (old id 3628221)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/23486562?dopt=Abstract
date added to LUP
2016-04-01 10:39:09
date last changed
2022-04-28 00:05:33
@article{77e282d8-1f81-4a5a-afe0-8210a7cbb999,
  abstract     = {{BackgroundTime-resolved three-dimensional, three-directional phase-contrast magnetic resonance velocity mapping (4D PC-MRI) is a powerful technique to depict dynamic blood flow patterns in the human body. However, the impact of phase background effects on flow visualizations has not been thoroughly studied previously, and it has not yet been experimentally demonstrated to what degree phase offsets affect flow visualizations and create errors such as inaccurate particle traces.PurposeTo quantify background phase offsets and their subsequent impact on particle trace visualizations in a 4D PC-MRI sequence. Additionally, we sought to investigate to what degree visualization errors are reduced by background phase correction.Material and MethodsA rotating phantom with a known velocity field was used to quantify background phase of 4D PC-MRI sequences accelerated with SENSE as well as different k-t BLAST speed-up factors. The deviation in end positions between particle traces in the measured velocity fields were compared before and after the application of two different phase correction methods.ResultsPhantom measurements revealed background velocity offsets up to 7 cm/s (7% of velocity encoding sensitivity) in the central slice, increasing with distance from the center. Background offsets remained constant with increasing k-t BLAST speed-up factors. End deviations of up to 5.3 mm (1.8 voxels) in the direction perpendicular to the rotating disc were found between particle traces and the seeding plane of the traces. Phase correction by subtraction of the data from the stationary phantom reduced the average deviation by up to 56%, while correcting the data-set with a first-order polynomial fit to stationary regions decreased average deviation up to 78%.ConclusionPathline visualizations can be significantly affected by background phase errors, highlighting the importance of dedicated and robust phase correction methods. Our results show that pathline deviation can be substantial if adequate phase background errors are not minimized.}},
  author       = {{Nilsson, Anders and Markenroth Bloch, Karin and Töger, Johannes and Heiberg, Einar and Ståhlberg, Freddy}},
  issn         = {{1600-0455}},
  language     = {{eng}},
  number       = {{6}},
  pages        = {{663--671}},
  publisher    = {{SAGE Publications}},
  series       = {{Acta Radiologica}},
  title        = {{Accuracy of four-dimensional phase-contrast velocity mapping for blood flow visualizations: a phantom study.}},
  url          = {{https://lup.lub.lu.se/search/files/2025864/3810099.pdf}},
  doi          = {{10.1177/0284185113478005}},
  volume       = {{54}},
  year         = {{2013}},
}