Abstract
In SPECT imaging of the chest, nonuniform attenuation

correction requires use of a patient specific attenuation (p)map. Such a map can be obtained by estimating the regions of 1)the lungs and 2) the soft tissues and bones, and then assigning an appropriate value of attenuation coefficient (p) to each region. We proposed a method to segment such regions from the Compton scatter and photopeak window SPECT slices of Tc-99m Sestamibi

studies. The Compton scatter slices are used to segment the body outline and to estimate the regions of the lungs. Locations of the back bone and sternum are estimated from the photopeak window slices to assist in the segmentation. To investigate the accuracy of using Compton scatter slices in... (More)
In SPECT imaging of the chest, nonuniform attenuation

correction requires use of a patient specific attenuation (p)map. Such a map can be obtained by estimating the regions of 1)the lungs and 2) the soft tissues and bones, and then assigning an appropriate value of attenuation coefficient (p) to each region. We proposed a method to segment such regions from the Compton scatter and photopeak window SPECT slices of Tc-99m Sestamibi

studies. The Compton scatter slices are used to segment the body outline and to estimate the regions of the lungs. Locations of the back bone and sternum are estimated from the photopeak window slices to assist in the segmentation. To investigate the accuracy of using Compton scatter slices in estimating the regions of the body and the lungs, a Monte-Carlo SPECT simulation of an anthropomorphic phantom with an activity distribution and noise characteristics similar to patient data was conducted. Energy windows of various widths were simulated for use in locating a suitable Compton scatter window for imaging. The effects of attenuation correction using a p map based on segmentation were also studied. The results demonstrated for the activity and p maps studied herein that: 1) reasonable contrast could be obtained from Compton scatter data for the segmentation of the lung regions, 2) true positive rates of 99% and 89% for

determining the body and lung regions, respectively, with total error rates of 4% and 29%, could be achieved, 3) usage of a p map based on segmentation for attenuation correction improved relative quantification over filtered backprojection, 4) variations in the assigned p value of 40% smaller or 40% larger in the lung regions had an insignificant impact on the results of relative quantification, 5) a wide energy window away from the photopeak window for recording scattered events could benefit both the segmentation of the lung regions and the attenuation correction of the activity in the myocardium region, and 6) usage of a smaller than true p value in the lung regions of an assigned p map might

benefit attenuation correction for absolute quantification. (Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
IEEE Transactions on Medical Imaging
volume
15
issue
1
pages
13 - 24
publisher
IEEE - Institute of Electrical and Electronics Engineers Inc.
ISSN
1558-254X
language
English
LU publication?
yes
id
7d49a7f5-81a9-408d-90ec-35703cc8eb95 (old id 1110560)
date added to LUP
2016-04-01 11:58:22
date last changed
2018-11-21 20:02:20
@article{7d49a7f5-81a9-408d-90ec-35703cc8eb95,
  abstract     = {{In SPECT imaging of the chest, nonuniform attenuation<br/><br>
correction requires use of a patient specific attenuation (p)map. Such a map can be obtained by estimating the regions of 1)the lungs and 2) the soft tissues and bones, and then assigning an appropriate value of attenuation coefficient (p) to each region. We proposed a method to segment such regions from the Compton scatter and photopeak window SPECT slices of Tc-99m Sestamibi<br/><br>
studies. The Compton scatter slices are used to segment the body outline and to estimate the regions of the lungs. Locations of the back bone and sternum are estimated from the photopeak window slices to assist in the segmentation. To investigate the accuracy of using Compton scatter slices in estimating the regions of the body and the lungs, a Monte-Carlo SPECT simulation of an anthropomorphic phantom with an activity distribution and noise characteristics similar to patient data was conducted. Energy windows of various widths were simulated for use in locating a suitable Compton scatter window for imaging. The effects of attenuation correction using a p map based on segmentation were also studied. The results demonstrated for the activity and p maps studied herein that: 1) reasonable contrast could be obtained from Compton scatter data for the segmentation of the lung regions, 2) true positive rates of 99% and 89% for<br/><br>
determining the body and lung regions, respectively, with total error rates of 4% and 29%, could be achieved, 3) usage of a p map based on segmentation for attenuation correction improved relative quantification over filtered backprojection, 4) variations in the assigned p value of 40% smaller or 40% larger in the lung regions had an insignificant impact on the results of relative quantification, 5) a wide energy window away from the photopeak window for recording scattered events could benefit both the segmentation of the lung regions and the attenuation correction of the activity in the myocardium region, and 6) usage of a smaller than true p value in the lung regions of an assigned p map might<br/><br>
benefit attenuation correction for absolute quantification.}},
  author       = {{Pan, Tin-Su and King, Michael A and DeVries, Daniel J and Ljungberg, Michael}},
  issn         = {{1558-254X}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{13--24}},
  publisher    = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}},
  series       = {{IEEE Transactions on Medical Imaging}},
  title        = {{Segmentation of the Body and Lungs from Compton Scatter and Photopeak Window Images in SPECT: A Monte Carlo Investigation}},
  volume       = {{15}},
  year         = {{1996}},
}