Scatter and crosstalk corrections for (99m)Tc/(123)I dual-radionuclide imaging using a CZT SPECT system with pinhole collimators.

Fan, Peng; Hutton, Brian F; Holstensson, Maria; Ljungberg, Michael LU ; Hendrik Pretorius, P; Prasad, Rameshwar; Ma, Tianyu; Liu, Yaqiang; Wang, Shi and Thorn, Stephanie L, et al. (2015) In Medical Physics 42(12). p.6895-6911
Abstract
The energy spectrum for a cadmium zinc telluride (CZT) detector has a low energy tail due to incomplete charge collection and intercrystal scattering. Due to these solid-state detector effects, scatter would be overestimated if the conventional triple-energy window (TEW) method is used for scatter and crosstalk corrections in CZT-based imaging systems. The objective of this work is to develop a scatter and crosstalk correction method for (99m)Tc/(123)I dual-radionuclide imaging for a CZT-based dedicated cardiac SPECT system with pinhole collimators (GE Discovery NM 530c/570c).
Please use this url to cite or link to this publication:
author
, et al. (More)
(Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Medical Physics
volume
42
issue
12
pages
6895 - 6911
publisher
American Association of Physicists in Medicine
external identifiers
  • pmid:26632046
  • wos:000368244100013
  • scopus:84947447609
ISSN
0094-2405
DOI
10.1118/1.4934830
language
English
LU publication?
yes
id
ff4cf8df-095f-436c-8e80-b102a0c42044 (old id 8505702)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/26632046?dopt=Abstract
date added to LUP
2016-01-04 21:34:09
date last changed
2017-06-04 03:51:54
@article{ff4cf8df-095f-436c-8e80-b102a0c42044,
  abstract     = {The energy spectrum for a cadmium zinc telluride (CZT) detector has a low energy tail due to incomplete charge collection and intercrystal scattering. Due to these solid-state detector effects, scatter would be overestimated if the conventional triple-energy window (TEW) method is used for scatter and crosstalk corrections in CZT-based imaging systems. The objective of this work is to develop a scatter and crosstalk correction method for (99m)Tc/(123)I dual-radionuclide imaging for a CZT-based dedicated cardiac SPECT system with pinhole collimators (GE Discovery NM 530c/570c).},
  author       = {Fan, Peng and Hutton, Brian F and Holstensson, Maria and Ljungberg, Michael and Hendrik Pretorius, P and Prasad, Rameshwar and Ma, Tianyu and Liu, Yaqiang and Wang, Shi and Thorn, Stephanie L and Stacy, Mitchel R and Sinusas, Albert J and Liu, Chi},
  issn         = {0094-2405},
  language     = {eng},
  number       = {12},
  pages        = {6895--6911},
  publisher    = {American Association of Physicists in Medicine},
  series       = {Medical Physics},
  title        = {Scatter and crosstalk corrections for (99m)Tc/(123)I dual-radionuclide imaging using a CZT SPECT system with pinhole collimators.},
  url          = {http://dx.doi.org/10.1118/1.4934830},
  volume       = {42},
  year         = {2015},
}