Scatter and crosstalk corrections for (99m)Tc/(123)I dual-radionuclide imaging using a CZT SPECT system with pinhole collimators.

Fan, Peng ; Hutton, Brian F ; Holstensson, Maria ; Ljungberg, Michael LU ; Hendrik Pretorius, P ; Prasad, Rameshwar ; Ma, Tianyu ; Liu, Yaqiang ; Wang, Shi and Thorn, Stephanie L , et al. (2015) In Medical Physics 42(12). p.6895-6911
Abstract
The energy spectrum for a cadmium zinc telluride (CZT) detector has a low energy tail due to incomplete charge collection and intercrystal scattering. Due to these solid-state detector effects, scatter would be overestimated if the conventional triple-energy window (TEW) method is used for scatter and crosstalk corrections in CZT-based imaging systems. The objective of this work is to develop a scatter and crosstalk correction method for (99m)Tc/(123)I dual-radionuclide imaging for a CZT-based dedicated cardiac SPECT system with pinhole collimators (GE Discovery NM 530c/570c).
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and , et al. (More)
; ; ; ; ; ; ; ; ; ; ; and (Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Medical Physics
volume
42
issue
12
pages
6895 - 6911
publisher
American Association of Physicists in Medicine
external identifiers
  • pmid:26632046
  • wos:000368244100013
  • scopus:84947447609
  • pmid:26632046
ISSN
0094-2405
DOI
10.1118/1.4934830
language
English
LU publication?
yes
id
ff4cf8df-095f-436c-8e80-b102a0c42044 (old id 8505702)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/26632046?dopt=Abstract
date added to LUP
2016-04-01 13:08:51
date last changed
2022-02-19 03:12:49
@article{ff4cf8df-095f-436c-8e80-b102a0c42044,
  abstract     = {{The energy spectrum for a cadmium zinc telluride (CZT) detector has a low energy tail due to incomplete charge collection and intercrystal scattering. Due to these solid-state detector effects, scatter would be overestimated if the conventional triple-energy window (TEW) method is used for scatter and crosstalk corrections in CZT-based imaging systems. The objective of this work is to develop a scatter and crosstalk correction method for (99m)Tc/(123)I dual-radionuclide imaging for a CZT-based dedicated cardiac SPECT system with pinhole collimators (GE Discovery NM 530c/570c).}},
  author       = {{Fan, Peng and Hutton, Brian F and Holstensson, Maria and Ljungberg, Michael and Hendrik Pretorius, P and Prasad, Rameshwar and Ma, Tianyu and Liu, Yaqiang and Wang, Shi and Thorn, Stephanie L and Stacy, Mitchel R and Sinusas, Albert J and Liu, Chi}},
  issn         = {{0094-2405}},
  language     = {{eng}},
  number       = {{12}},
  pages        = {{6895--6911}},
  publisher    = {{American Association of Physicists in Medicine}},
  series       = {{Medical Physics}},
  title        = {{Scatter and crosstalk corrections for (99m)Tc/(123)I dual-radionuclide imaging using a CZT SPECT system with pinhole collimators.}},
  url          = {{http://dx.doi.org/10.1118/1.4934830}},
  doi          = {{10.1118/1.4934830}},
  volume       = {{42}},
  year         = {{2015}},
}