Multicenter study for the evaluation of a dedicated biopsy device for MR-guided vacuum biopsy of the breast
(2002) In European Radiology 12(6). p.1463-1470- Abstract
- The purpose of this multicenter study was to determine the accuracy and clinical value of a dedicated breast biopsy system which allows for MR-guided vacuum biopsy (VB) of contrast-enhancing lesions. In five European centers, MR-guided 11-gauge VB was performed on 341 lesions. In 7 cases VB was unsuccessful. This was immediately realized on postinterventional images or direct follow-up combined with histopathology-imaging correlation; thus, a false-negative diagnosis was avoided. Histology of 334 successful biopsies yielded 84 (25%) malignancies, 17 (5%) atypical ductal hyperplasias, and 233 (70%) benign entities. Verification of malignant or borderline lesions included re-excision of the biopsy cavity. Benign histologic biopsy results... (More)
- The purpose of this multicenter study was to determine the accuracy and clinical value of a dedicated breast biopsy system which allows for MR-guided vacuum biopsy (VB) of contrast-enhancing lesions. In five European centers, MR-guided 11-gauge VB was performed on 341 lesions. In 7 cases VB was unsuccessful. This was immediately realized on postinterventional images or direct follow-up combined with histopathology-imaging correlation; thus, a false-negative diagnosis was avoided. Histology of 334 successful biopsies yielded 84 (25%) malignancies, 17 (5%) atypical ductal hyperplasias, and 233 (70%) benign entities. Verification of malignant or borderline lesions included re-excision of the biopsy cavity. Benign histologic biopsy results were verified by retrospective correlation with the pre- and postinterventional MRI and by subsequent follow-up. Our results indicate that MR-guided VB, in combination with the dedicated biopsy coil, offers the possibility to accurately diagnose even very small lesions that can only be visualized or localized by MRI. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/893119
- author
- organization
- publishing date
- 2002
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- intervention, biopsy, breast, MR imaging
- in
- European Radiology
- volume
- 12
- issue
- 6
- pages
- 1463 - 1470
- publisher
- Springer
- external identifiers
-
- wos:000176197400028
- pmid:12042955
- scopus:0036598381
- pmid:12042955
- ISSN
- 0938-7994
- DOI
- 10.1007/s00330-002-1376-4
- language
- English
- LU publication?
- yes
- id
- 9effb180-9b50-406a-b09c-524ac780e5d8 (old id 893119)
- date added to LUP
- 2016-04-01 12:22:12
- date last changed
- 2022-01-27 02:48:41
@article{9effb180-9b50-406a-b09c-524ac780e5d8, abstract = {{The purpose of this multicenter study was to determine the accuracy and clinical value of a dedicated breast biopsy system which allows for MR-guided vacuum biopsy (VB) of contrast-enhancing lesions. In five European centers, MR-guided 11-gauge VB was performed on 341 lesions. In 7 cases VB was unsuccessful. This was immediately realized on postinterventional images or direct follow-up combined with histopathology-imaging correlation; thus, a false-negative diagnosis was avoided. Histology of 334 successful biopsies yielded 84 (25%) malignancies, 17 (5%) atypical ductal hyperplasias, and 233 (70%) benign entities. Verification of malignant or borderline lesions included re-excision of the biopsy cavity. Benign histologic biopsy results were verified by retrospective correlation with the pre- and postinterventional MRI and by subsequent follow-up. Our results indicate that MR-guided VB, in combination with the dedicated biopsy coil, offers the possibility to accurately diagnose even very small lesions that can only be visualized or localized by MRI.}}, author = {{Perlet, C and Heinig, A and Prat, X and Casselman, J and Bååth, Lars and Sittek, H and Stets, C and Lamarque, J and Anderson, I and Schneider, P and Taourel, P and Reiser, M and Heywang-Kobrunner, SH}}, issn = {{0938-7994}}, keywords = {{intervention; biopsy; breast; MR imaging}}, language = {{eng}}, number = {{6}}, pages = {{1463--1470}}, publisher = {{Springer}}, series = {{European Radiology}}, title = {{Multicenter study for the evaluation of a dedicated biopsy device for MR-guided vacuum biopsy of the breast}}, url = {{http://dx.doi.org/10.1007/s00330-002-1376-4}}, doi = {{10.1007/s00330-002-1376-4}}, volume = {{12}}, year = {{2002}}, }