Sensitivity enhancement and fringe reduction in tunable diode laser spectroscopy using hemispherical diffusers
(2017) In Review of Scientific Instruments 88(5).- Abstract
The use of diffuse, highly reflective optical components, in particular, a hemispherical BaSO4 diffuser, at the point of light injection into non-transparent or turbid media was evaluated as a means to increase the measurement sensitivity of spectroscopic absorption measurements. By performing the light injection from, e.g., an optical fiber through a component designed to make the light diffuse and to reflect (and thereby re-inject) light scattered from the sample, the total amount of light delivered into the sample is increased. Further, the occurrence of possible interference fringes is strongly reduced.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/9652bfbc-a12a-463d-93bb-5213daf6c341
- author
- Lundin, Patrik LU ; Karlsson, Marcus ; Mei, Liang LU ; Larsson, Jim LU ; Somesfalean, Gabriel LU and Svanberg, Sune LU
- organization
- publishing date
- 2017-05-01
- type
- Contribution to journal
- publication status
- published
- subject
- in
- Review of Scientific Instruments
- volume
- 88
- issue
- 5
- article number
- 053111
- publisher
- American Institute of Physics (AIP)
- external identifiers
-
- pmid:28571416
- wos:000402801900021
- scopus:85019708005
- ISSN
- 0034-6748
- DOI
- 10.1063/1.4983807
- language
- English
- LU publication?
- yes
- id
- 9652bfbc-a12a-463d-93bb-5213daf6c341
- date added to LUP
- 2017-06-13 09:33:33
- date last changed
- 2024-10-14 07:45:07
@article{9652bfbc-a12a-463d-93bb-5213daf6c341, abstract = {{<p>The use of diffuse, highly reflective optical components, in particular, a hemispherical BaSO<sub>4</sub> diffuser, at the point of light injection into non-transparent or turbid media was evaluated as a means to increase the measurement sensitivity of spectroscopic absorption measurements. By performing the light injection from, e.g., an optical fiber through a component designed to make the light diffuse and to reflect (and thereby re-inject) light scattered from the sample, the total amount of light delivered into the sample is increased. Further, the occurrence of possible interference fringes is strongly reduced.</p>}}, author = {{Lundin, Patrik and Karlsson, Marcus and Mei, Liang and Larsson, Jim and Somesfalean, Gabriel and Svanberg, Sune}}, issn = {{0034-6748}}, language = {{eng}}, month = {{05}}, number = {{5}}, publisher = {{American Institute of Physics (AIP)}}, series = {{Review of Scientific Instruments}}, title = {{Sensitivity enhancement and fringe reduction in tunable diode laser spectroscopy using hemispherical diffusers}}, url = {{http://dx.doi.org/10.1063/1.4983807}}, doi = {{10.1063/1.4983807}}, volume = {{88}}, year = {{2017}}, }