Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

Multiple window spectrogram of peaks due to transients in the electroencephalogram

Sandsten, Maria LU and Lindgren, Magnus LU (2001) In IEEE Transactions on Biomedical Engineering 48(3). p.284-293
Abstract
Peak matched multiple windows (PM MW) were used to estimate the spectrogram of the electroencephalogram (EEG). The authors focussed on the ability to estimate frequency changes, and especially resolving close peaks. A peak of known frequency was evoked in the EEG in a predetermined time interval. The PM MW spectrogram was compared to the commonly used single Hanning window and to weighted overlapped segment averaging in simulations and for real-data. The PM MW were shown to give estimates with good resolution and low variance
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
good resolution low variance estimates, electroencephalogram peaks, EEG transients, electrodiagnostics, spectral analysis, electroencephalography, medical signal processing, single Hanning window, weighted overlapped segment averaging, multiple-window spectrogram
in
IEEE Transactions on Biomedical Engineering
volume
48
issue
3
pages
284 - 293
publisher
IEEE - Institute of Electrical and Electronics Engineers Inc.
external identifiers
  • scopus:0035096537
ISSN
1558-2531
DOI
10.1109/10.914791
language
English
LU publication?
yes
id
de254eba-5e10-471e-88b9-56816e7b6b77 (old id 995844)
date added to LUP
2016-04-04 09:22:47
date last changed
2022-01-29 17:33:07
@article{de254eba-5e10-471e-88b9-56816e7b6b77,
  abstract     = {{Peak matched multiple windows (PM MW) were used to estimate the spectrogram of the electroencephalogram (EEG). The authors focussed on the ability to estimate frequency changes, and especially resolving close peaks. A peak of known frequency was evoked in the EEG in a predetermined time interval. The PM MW spectrogram was compared to the commonly used single Hanning window and to weighted overlapped segment averaging in simulations and for real-data. The PM MW were shown to give estimates with good resolution and low variance}},
  author       = {{Sandsten, Maria and Lindgren, Magnus}},
  issn         = {{1558-2531}},
  keywords     = {{good resolution low variance estimates; electroencephalogram peaks; EEG transients; electrodiagnostics; spectral analysis; electroencephalography; medical signal processing; single Hanning window; weighted overlapped segment averaging; multiple-window spectrogram}},
  language     = {{eng}},
  number       = {{3}},
  pages        = {{284--293}},
  publisher    = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}},
  series       = {{IEEE Transactions on Biomedical Engineering}},
  title        = {{Multiple window spectrogram of peaks due to transients in the electroencephalogram}},
  url          = {{http://dx.doi.org/10.1109/10.914791}},
  doi          = {{10.1109/10.914791}},
  volume       = {{48}},
  year         = {{2001}},
}