Pareto optimization of SPECT acquisition and reconstruction settings for 177Lu activity quantification

Gustafsson, Johan LU ; Larsson, Erik LU ; Ljungberg, Michael LU and Sjögreen Gleisner, Katarina LU (2024) In EJNMMI Physics 11(1).
Abstract

Background: The aim was to investigate the noise and bias properties of quantitative 177Lu-SPECT with respect to the number of projection angles, and the number of subsets and iterations in the OS-EM reconstruction, for different total acquisition times. Methods: Experimental SPECT acquisition of six spheres in a NEMA body phantom filled with 177Lu was performed, using medium-energy collimators and 120 projections with 180 s per projection. Bootstrapping was applied to generate data sets representing acquisitions with 20 to 120 projections for 10 min, 20 min, and 40 min, with 32 noise realizations per setting. Monte Carlo simulations were performed of 177Lu-DOTA-TATE in an anthropomorphic computer... (More)

Background: The aim was to investigate the noise and bias properties of quantitative 177Lu-SPECT with respect to the number of projection angles, and the number of subsets and iterations in the OS-EM reconstruction, for different total acquisition times. Methods: Experimental SPECT acquisition of six spheres in a NEMA body phantom filled with 177Lu was performed, using medium-energy collimators and 120 projections with 180 s per projection. Bootstrapping was applied to generate data sets representing acquisitions with 20 to 120 projections for 10 min, 20 min, and 40 min, with 32 noise realizations per setting. Monte Carlo simulations were performed of 177Lu-DOTA-TATE in an anthropomorphic computer phantom with three tumours (2.8 mL to 40.0 mL). Projections representing 24 h and 168 h post administration were simulated, each with 32 noise realizations. Images were reconstructed using OS-EM with compensation for attenuation, scatter, and distance-dependent resolution. The number of subsets and iterations were varied within a constrained range of the product number of iterations × number of projections ≤2400. Volumes-of-interest were defined following the physical size of the spheres and tumours, the mean activity-concentrations estimated, and the absolute mean relative error and coefficient of variation (CV) over noise realizations calculated. Pareto fronts were established by analysis of CV versus mean relative error. Results: Points at the Pareto fronts with low CV and high mean error resulted from using a low number of subsets, whilst points at the Pareto fronts associated with high CV but low mean error resulted from reconstructions with a high number of subsets. The number of projection angles had limited impact. Conclusions: For accurate estimation of the 177Lu activity-concentration from SPECT images, the number of projection angles has limited importance, whilst the total acquisition time and the number of subsets and iterations are parameters of importance.

(Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Lu, Pareto optimization, Quantitative SPECT, Reconstruction
in
EJNMMI Physics
volume
11
issue
1
article number
62
publisher
Springer
external identifiers
  • scopus:85198372726
ISSN
2197-7364
DOI
10.1186/s40658-024-00667-7
language
English
LU publication?
yes
id
99b94ac8-d349-410c-99e4-c7e4138f2644
date added to LUP
2024-08-30 11:35:19
date last changed
2024-09-24 08:03:49
@article{99b94ac8-d349-410c-99e4-c7e4138f2644,
  abstract     = {{<p>Background: The aim was to investigate the noise and bias properties of quantitative <sup>177</sup>Lu-SPECT with respect to the number of projection angles, and the number of subsets and iterations in the OS-EM reconstruction, for different total acquisition times. Methods: Experimental SPECT acquisition of six spheres in a NEMA body phantom filled with <sup>177</sup>Lu was performed, using medium-energy collimators and 120 projections with 180 s per projection. Bootstrapping was applied to generate data sets representing acquisitions with 20 to 120 projections for 10 min, 20 min, and 40 min, with 32 noise realizations per setting. Monte Carlo simulations were performed of <sup>177</sup>Lu-DOTA-TATE in an anthropomorphic computer phantom with three tumours (2.8 mL to 40.0 mL). Projections representing 24 h and 168 h post administration were simulated, each with 32 noise realizations. Images were reconstructed using OS-EM with compensation for attenuation, scatter, and distance-dependent resolution. The number of subsets and iterations were varied within a constrained range of the product number of iterations × number of projections ≤2400. Volumes-of-interest were defined following the physical size of the spheres and tumours, the mean activity-concentrations estimated, and the absolute mean relative error and coefficient of variation (CV) over noise realizations calculated. Pareto fronts were established by analysis of CV versus mean relative error. Results: Points at the Pareto fronts with low CV and high mean error resulted from using a low number of subsets, whilst points at the Pareto fronts associated with high CV but low mean error resulted from reconstructions with a high number of subsets. The number of projection angles had limited impact. Conclusions: For accurate estimation of the <sup>177</sup>Lu activity-concentration from SPECT images, the number of projection angles has limited importance, whilst the total acquisition time and the number of subsets and iterations are parameters of importance.</p>}},
  author       = {{Gustafsson, Johan and Larsson, Erik and Ljungberg, Michael and Sjögreen Gleisner, Katarina}},
  issn         = {{2197-7364}},
  keywords     = {{Lu; Pareto optimization; Quantitative SPECT; Reconstruction}},
  language     = {{eng}},
  number       = {{1}},
  publisher    = {{Springer}},
  series       = {{EJNMMI Physics}},
  title        = {{Pareto optimization of SPECT acquisition and reconstruction settings for <sup>177</sup>Lu activity quantification}},
  url          = {{http://dx.doi.org/10.1186/s40658-024-00667-7}},
  doi          = {{10.1186/s40658-024-00667-7}},
  volume       = {{11}},
  year         = {{2024}},
}