Performance assessment of photon migration instruments: the MEDPHOT protocol

Pifferi, A ; Torricelli, A ; Bassi, A ; Taroni, P ; Cubeddu, R ; Wabnitz, H ; Grosenick, D ; Moller, M ; Macdonald, R and Swartling, Johannes LU , et al. (2005) In Applied Optics 44(11). p.2104-2114
Abstract
We propose a comprehensive protocol for the performance assessment of photon migration instruments. The protocol has been developed within the European Thematic Network MEDPHOT (optical methods for medical diagnosis and monitoring of diseases) and is based on five criteria: accuracy, linearity, noise, stability, and reproducibility. This protocol was applied to a total of 8 instruments with a set of 32 phantoms, covering a wide range of optical properties.
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and , et al. (More)
; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; and (Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Applied Optics
volume
44
issue
11
pages
2104 - 2114
publisher
Optical Society of America
external identifiers
  • wos:000228457800013
  • pmid:15838951
  • scopus:20244374548
ISSN
2155-3165
language
English
LU publication?
yes
id
9a93ec4d-9c2e-45d1-a69c-76b901c3e7ae (old id 245845)
alternative location
http://www.opticsinfobase.org/DirectPDFAccess/8DC89F06-BDB9-137E-C01C0BE59752FCC1_83248.pdf?da=1&id=83248&seq=0&CFID=47847101&CFTOKEN=13567969
date added to LUP
2016-04-01 12:14:17
date last changed
2022-04-29 02:28:55
@article{9a93ec4d-9c2e-45d1-a69c-76b901c3e7ae,
  abstract     = {{We propose a comprehensive protocol for the performance assessment of photon migration instruments. The protocol has been developed within the European Thematic Network MEDPHOT (optical methods for medical diagnosis and monitoring of diseases) and is based on five criteria: accuracy, linearity, noise, stability, and reproducibility. This protocol was applied to a total of 8 instruments with a set of 32 phantoms, covering a wide range of optical properties.}},
  author       = {{Pifferi, A and Torricelli, A and Bassi, A and Taroni, P and Cubeddu, R and Wabnitz, H and Grosenick, D and Moller, M and Macdonald, R and Swartling, Johannes and Svensson, Tomas and Andersson-Engels, Stefan and van Veen, RLP and Sterenborg, HJCM and Tualle, JM and Nghiem, HL and Avrillier, S and Whelan, M and Stamm, H}},
  issn         = {{2155-3165}},
  language     = {{eng}},
  number       = {{11}},
  pages        = {{2104--2114}},
  publisher    = {{Optical Society of America}},
  series       = {{Applied Optics}},
  title        = {{Performance assessment of photon migration instruments: the MEDPHOT protocol}},
  url          = {{https://lup.lub.lu.se/search/files/2840019/2370740.pdf}},
  volume       = {{44}},
  year         = {{2005}},
}