Impact of DOI in a Clinical SPECT/MRI System : A Simulation Study

Morahan, Ashley J. ; Erlandsson, Kjell ; Cerrato, Annalisa ; D'Adda, Ilenia ; Carminati, Marco ; Ljungberg, Michael LU ; Fiorini, Carlo E. and Hutton, Brian F. (2020) 2020 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2020
Abstract

A novel SPECT/MRI scanner has been modelled and tested here using Monte Carlo simulation software, SIMIND. The INSERT SPECT/MRI system faces challenges with event reconstruction due to photon depth of interaction. The novel SPECT system is subject to parallax errors due to its crystal size and slit aperture collimator. We present a simple measure of the DOI errors through SIMIND experiments; by modelling the DOI layers we are able to improve the reconstruction of projection data in the INSERT scanner. A set of capillary phantoms are simulated to explore the impact of DOI on the resolution of the scanner and establish corrections in the system's reconstruction.

Please use this url to cite or link to this publication:
author
; ; ; ; ; ; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
host publication
2020 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2020
publisher
IEEE - Institute of Electrical and Electronics Engineers Inc.
conference name
2020 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2020
conference location
Boston, United States
conference dates
2020-10-31 - 2020-11-07
external identifiers
  • scopus:85124705182
ISBN
9781728176932
DOI
10.1109/NSS/MIC42677.2020.9508027
language
English
LU publication?
yes
id
a005ace0-0fd7-431e-b2ea-5abc4d92b8b8
date added to LUP
2022-04-01 15:43:33
date last changed
2022-04-24 23:57:34
@inproceedings{a005ace0-0fd7-431e-b2ea-5abc4d92b8b8,
  abstract     = {{<p>A novel SPECT/MRI scanner has been modelled and tested here using Monte Carlo simulation software, SIMIND. The INSERT SPECT/MRI system faces challenges with event reconstruction due to photon depth of interaction. The novel SPECT system is subject to parallax errors due to its crystal size and slit aperture collimator. We present a simple measure of the DOI errors through SIMIND experiments; by modelling the DOI layers we are able to improve the reconstruction of projection data in the INSERT scanner. A set of capillary phantoms are simulated to explore the impact of DOI on the resolution of the scanner and establish corrections in the system's reconstruction.</p>}},
  author       = {{Morahan, Ashley J. and Erlandsson, Kjell and Cerrato, Annalisa and D'Adda, Ilenia and Carminati, Marco and Ljungberg, Michael and Fiorini, Carlo E. and Hutton, Brian F.}},
  booktitle    = {{2020 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2020}},
  isbn         = {{9781728176932}},
  language     = {{eng}},
  publisher    = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}},
  title        = {{Impact of DOI in a Clinical SPECT/MRI System : A Simulation Study}},
  url          = {{http://dx.doi.org/10.1109/NSS/MIC42677.2020.9508027}},
  doi          = {{10.1109/NSS/MIC42677.2020.9508027}},
  year         = {{2020}},
}