Pencil beam scanning proton therapy for mediastinal lymphomas in deep inspiration breath-hold : a retrospective assessment of plan robustness

Hörberger, Filip LU ; Andersson, Karin M. ; Enmark, Marika ; Kristensen, Ingrid LU ; Flejmer, Anna and Edvardsson, Anneli LU (2024) In Acta oncologica (Stockholm, Sweden) 63. p.62-69
Abstract

PURPOSE/BACKGROUND: The aim of this study was to evaluate pencil beam scanning (PBS) proton therapy (PT) in deep inspiration breath-hold (DIBH) for mediastinal lymphoma patients, by retrospectively evaluating plan robustness to the clinical target volume (CTV) and organs at risk (OARs) on repeated CT images acquired throughout treatment.  Methods: Sixteen mediastinal lymphoma patients treated with PBS-PT in DIBH were included. Treatment plans (TPs) were robustly optimized on the CTV (7 mm/4.5%). Repeated verification CTs (vCT) were acquired during the treatment course, resulting in 52 images for the entire patient cohort. The CTV and OARs were transferred from the planning CT to the vCTs with deformable image registration and the TPs... (More)

PURPOSE/BACKGROUND: The aim of this study was to evaluate pencil beam scanning (PBS) proton therapy (PT) in deep inspiration breath-hold (DIBH) for mediastinal lymphoma patients, by retrospectively evaluating plan robustness to the clinical target volume (CTV) and organs at risk (OARs) on repeated CT images acquired throughout treatment.  Methods: Sixteen mediastinal lymphoma patients treated with PBS-PT in DIBH were included. Treatment plans (TPs) were robustly optimized on the CTV (7 mm/4.5%). Repeated verification CTs (vCT) were acquired during the treatment course, resulting in 52 images for the entire patient cohort. The CTV and OARs were transferred from the planning CT to the vCTs with deformable image registration and the TPs were recalculated on the vCTs. Target coverage and OAR doses at the vCTs were compared to the nominal plan. Deviation in lung volume was also calculated. RESULTS: The TPs demonstrated high robust target coverage throughout treatment with D98%,CTV deviations within 2% for 14 patients and above the desired requirement of 95% for 49/52 vCTs. However, two patients did not achieve a robust dose to CTV due to poor DIBH reproducibility, with D98%,CTV at 78 and 93% respectively, and replanning was performed for one patient. Adequate OAR sparing was achieved for all patients. Total lung volume variation was below 10% for 39/52 vCTs. CONCLUSION: PBS PT in DIBH is generally a robust technique for treatment of mediastinal lymphomas. However, closely monitoring the DIBH-reproducibility during treatment is important to avoid underdosing CTV and achieve sufficient dose-sparing of the OARs.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Acta oncologica (Stockholm, Sweden)
volume
63
pages
8 pages
publisher
Taylor & Francis
external identifiers
  • pmid:38415848
  • scopus:85186740985
ISSN
1651-226X
DOI
10.2340/1651-226X.2024.23964
language
English
LU publication?
yes
id
a030d970-bb23-4a82-9208-82db3ed655ab
date added to LUP
2024-03-21 15:59:49
date last changed
2024-04-18 16:28:24
@article{a030d970-bb23-4a82-9208-82db3ed655ab,
  abstract     = {{<p>PURPOSE/BACKGROUND: The aim of this study was to evaluate pencil beam scanning (PBS) proton therapy (PT) in deep inspiration breath-hold (DIBH) for mediastinal lymphoma patients, by retrospectively evaluating plan robustness to the clinical target volume (CTV) and organs at risk (OARs) on repeated CT images acquired throughout treatment.  Methods: Sixteen mediastinal lymphoma patients treated with PBS-PT in DIBH were included. Treatment plans (TPs) were robustly optimized on the CTV (7 mm/4.5%). Repeated verification CTs (vCT) were acquired during the treatment course, resulting in 52 images for the entire patient cohort. The CTV and OARs were transferred from the planning CT to the vCTs with deformable image registration and the TPs were recalculated on the vCTs. Target coverage and OAR doses at the vCTs were compared to the nominal plan. Deviation in lung volume was also calculated. RESULTS: The TPs demonstrated high robust target coverage throughout treatment with D98%,CTV deviations within 2% for 14 patients and above the desired requirement of 95% for 49/52 vCTs. However, two patients did not achieve a robust dose to CTV due to poor DIBH reproducibility, with D98%,CTV at 78 and 93% respectively, and replanning was performed for one patient. Adequate OAR sparing was achieved for all patients. Total lung volume variation was below 10% for 39/52 vCTs. CONCLUSION: PBS PT in DIBH is generally a robust technique for treatment of mediastinal lymphomas. However, closely monitoring the DIBH-reproducibility during treatment is important to avoid underdosing CTV and achieve sufficient dose-sparing of the OARs.</p>}},
  author       = {{Hörberger, Filip and Andersson, Karin M. and Enmark, Marika and Kristensen, Ingrid and Flejmer, Anna and Edvardsson, Anneli}},
  issn         = {{1651-226X}},
  language     = {{eng}},
  month        = {{02}},
  pages        = {{62--69}},
  publisher    = {{Taylor & Francis}},
  series       = {{Acta oncologica (Stockholm, Sweden)}},
  title        = {{Pencil beam scanning proton therapy for mediastinal lymphomas in deep inspiration breath-hold : a retrospective assessment of plan robustness}},
  url          = {{http://dx.doi.org/10.2340/1651-226X.2024.23964}},
  doi          = {{10.2340/1651-226X.2024.23964}},
  volume       = {{63}},
  year         = {{2024}},
}