A new collimator simulation in SIMIND based on the delta-scattering technique
(2005) In IEEE Transactions on Nuclear Science 52(5). p.1370-1375- Abstract
- To use conventional ray tracing methods in Monte Carlo simulation of the collimator in a scintillation camera system can be time consuming. It is however necessary to take collimator interactions into account when simulating radionuclides emitting high-energy photons that can penetrate the septa in the collimator. In this work a statistical collimator algorithm, based on the Delta-Scattering method, is evaluated using I-123. The evaluation is performed by comparing results from Monte Carlo simulations and measurements for a scintillation camera system, using point sources and a nonhomogeneous brain phantom. A good agreement can be seen for both images and energy spectra.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/211158
- author
- Ljungberg, Michael
LU
; Larsson, A
and Johansson, L
- organization
- publishing date
- 2005
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- SPECT, simulation, scattering, collimator, Monte Carlo
- in
- IEEE Transactions on Nuclear Science
- volume
- 52
- issue
- 5
- pages
- 1370 - 1375
- publisher
- IEEE - Institute of Electrical and Electronics Engineers Inc.
- external identifiers
-
- wos:000233780800024
- scopus:29144471251
- ISSN
- 0018-9499
- DOI
- 10.1109/TNS.2005.858252
- language
- English
- LU publication?
- yes
- id
- ba594f8f-5eb9-454c-84d9-e22d8f53e51c (old id 211158)
- date added to LUP
- 2016-04-01 16:06:08
- date last changed
- 2025-10-14 08:55:21
@article{ba594f8f-5eb9-454c-84d9-e22d8f53e51c,
abstract = {{To use conventional ray tracing methods in Monte Carlo simulation of the collimator in a scintillation camera system can be time consuming. It is however necessary to take collimator interactions into account when simulating radionuclides emitting high-energy photons that can penetrate the septa in the collimator. In this work a statistical collimator algorithm, based on the Delta-Scattering method, is evaluated using I-123. The evaluation is performed by comparing results from Monte Carlo simulations and measurements for a scintillation camera system, using point sources and a nonhomogeneous brain phantom. A good agreement can be seen for both images and energy spectra.}},
author = {{Ljungberg, Michael and Larsson, A and Johansson, L}},
issn = {{0018-9499}},
keywords = {{SPECT; simulation; scattering; collimator; Monte Carlo}},
language = {{eng}},
number = {{5}},
pages = {{1370--1375}},
publisher = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}},
series = {{IEEE Transactions on Nuclear Science}},
title = {{A new collimator simulation in SIMIND based on the delta-scattering technique}},
url = {{http://dx.doi.org/10.1109/TNS.2005.858252}},
doi = {{10.1109/TNS.2005.858252}},
volume = {{52}},
year = {{2005}},
}