Performance of a Personal Low-Weight Aerosol Sampler with Particle Size Fractionation and Time Resolution

Malmqvist, Klas LU ; Bohgard, Mats LU ; Johansson, Gerd LU and Akselsson, Roland LU (1983) The Tenth Annual Conference of the Association for Aerosol Research 14(3). p.409-413
Abstract
A low-weight aerosol sampler has been developed and tested. Internal losses for both sub-micron and

large particles ( < 10 tam) are of the order of 5 %. The size fractionation patterns indicate that the sampler may be

used to collect a separate respirable fraction in good agreement with the BMRC standard and a coarse fraction

deternuned by the inlet design.
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
keywords
aerosol sampler, size fractionation, time resolution
host publication
Journal of Aerosol Science
volume
14
issue
3
pages
409 - 413
conference name
The Tenth Annual Conference of the Association for Aerosol Research
conference location
Bologna, Italy
conference dates
1982-10-14 - 1982-10-17
external identifiers
  • scopus:0020543615
DOI
10.1016/0021-8502(83)90149-0
language
English
LU publication?
yes
additional info
The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Ergonomics and Aerosol Technology (011025002), Division of Occupational and Environmental Medicine (013078001), Nuclear Physics (Faculty of Technology) (011013007)
id
bc844a52-f884-464e-addd-46093c5a3cec (old id 1762785)
date added to LUP
2016-04-04 13:34:06
date last changed
2021-01-03 10:34:16
@inproceedings{bc844a52-f884-464e-addd-46093c5a3cec,
  abstract     = {{A low-weight aerosol sampler has been developed and tested. Internal losses for both sub-micron and<br/><br>
large particles ( &lt; 10 tam) are of the order of 5 %. The size fractionation patterns indicate that the sampler may be<br/><br>
used to collect a separate respirable fraction in good agreement with the BMRC standard and a coarse fraction<br/><br>
deternuned by the inlet design.}},
  author       = {{Malmqvist, Klas and Bohgard, Mats and Johansson, Gerd and Akselsson, Roland}},
  booktitle    = {{Journal of Aerosol Science}},
  keywords     = {{aerosol sampler; size fractionation; time resolution}},
  language     = {{eng}},
  number       = {{3}},
  pages        = {{409--413}},
  title        = {{Performance of a Personal Low-Weight Aerosol Sampler with Particle Size Fractionation and Time Resolution}},
  url          = {{http://dx.doi.org/10.1016/0021-8502(83)90149-0}},
  doi          = {{10.1016/0021-8502(83)90149-0}},
  volume       = {{14}},
  year         = {{1983}},
}