Nodule detection in digital chest radiography: Summary of the radius chest trial

Hakansson, M ; Bath, M ; Borjesson, S ; Kheddache, S ; Grahn, Anna LU ; Ruschin, Mark LU ; Tingberg, Anders LU ; Mattsson, Sören LU and Mansson, LG (2005) In Radiation Protection Dosimetry 114(1-3). p.114-120
Abstract
As a part of the Europe-wide research project 'Unification of physical and clinical requirements for medical X-ray imaging'-governed by the Radiological Imaging Unification Strategies (RADIUS) Group-a major image quality trial was conducted by members of the group. The RADIUS chest trial aimed at thoroughly examining various aspects of nodule detection in digital chest radiography, such as the effects of nodule location, system noise, anatomical noise, and anatomical background. The main findings of the RADIUS chest trial concerning the detection of a lung nodule with a size in the order of 10 mm can be summarised as: (1) the detectability of the nodule is largely dependent on its location in the chest. (2) the system noise has a minor... (More)
As a part of the Europe-wide research project 'Unification of physical and clinical requirements for medical X-ray imaging'-governed by the Radiological Imaging Unification Strategies (RADIUS) Group-a major image quality trial was conducted by members of the group. The RADIUS chest trial aimed at thoroughly examining various aspects of nodule detection in digital chest radiography, such as the effects of nodule location, system noise, anatomical noise, and anatomical background. The main findings of the RADIUS chest trial concerning the detection of a lung nodule with a size in the order of 10 mm can be summarised as: (1) the detectability of the nodule is largely dependent on its location in the chest. (2) the system noise has a minor impact on the detectability at the dose levels used today, (3) the disturbance of the anatomical noise is larger than that of the system noise but smaller than that of the anatomical background and (4) the anatomical background acts as noise to a large extent and is the major image component affecting the detectability of the nodule. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Radiation Protection Dosimetry
volume
114
issue
1-3
pages
114 - 120
publisher
Oxford University Press
external identifiers
  • wos:000229927400018
  • pmid:15933091
  • scopus:21244492363
ISSN
1742-3406
DOI
10.1093/rpd/nch574
language
English
LU publication?
yes
id
c08f5c71-e49e-4d6a-b28b-94472714d609 (old id 895212)
date added to LUP
2016-04-01 16:33:13
date last changed
2022-01-28 20:28:38
@article{c08f5c71-e49e-4d6a-b28b-94472714d609,
  abstract     = {{As a part of the Europe-wide research project 'Unification of physical and clinical requirements for medical X-ray imaging'-governed by the Radiological Imaging Unification Strategies (RADIUS) Group-a major image quality trial was conducted by members of the group. The RADIUS chest trial aimed at thoroughly examining various aspects of nodule detection in digital chest radiography, such as the effects of nodule location, system noise, anatomical noise, and anatomical background. The main findings of the RADIUS chest trial concerning the detection of a lung nodule with a size in the order of 10 mm can be summarised as: (1) the detectability of the nodule is largely dependent on its location in the chest. (2) the system noise has a minor impact on the detectability at the dose levels used today, (3) the disturbance of the anatomical noise is larger than that of the system noise but smaller than that of the anatomical background and (4) the anatomical background acts as noise to a large extent and is the major image component affecting the detectability of the nodule.}},
  author       = {{Hakansson, M and Bath, M and Borjesson, S and Kheddache, S and Grahn, Anna and Ruschin, Mark and Tingberg, Anders and Mattsson, Sören and Mansson, LG}},
  issn         = {{1742-3406}},
  language     = {{eng}},
  number       = {{1-3}},
  pages        = {{114--120}},
  publisher    = {{Oxford University Press}},
  series       = {{Radiation Protection Dosimetry}},
  title        = {{Nodule detection in digital chest radiography: Summary of the radius chest trial}},
  url          = {{http://dx.doi.org/10.1093/rpd/nch574}},
  doi          = {{10.1093/rpd/nch574}},
  volume       = {{114}},
  year         = {{2005}},
}