Accurate heat capacity determination of solids and liquids using a heat conduction calorimeter

Hothar, Marcus and Wadsö, Lars LU (2024) In Journal of Thermal Analysis and Calorimetry 149(5). p.2179-2188
Abstract

This study investigates the suitability of heat conduction calorimeters for determining the specific heat capacity of solid and liquid substances. Accurate and precise measurements were obtained for various substances, including water, ethylene glycol, the ionic liquid [EMIM][TCM], and copper, with relative standard deviations averaging less than 1%. Measurements on [EMIM][DCA] indicated a systematic deviation from the literature values. The study highlights the repeatability of the measurement method, which worked well for both temperature increases and decreases, as well as for the mean of the two. The influence of sample size on the results when it came to liquids was also investigated, revealing that large sample sizes led to... (More)

This study investigates the suitability of heat conduction calorimeters for determining the specific heat capacity of solid and liquid substances. Accurate and precise measurements were obtained for various substances, including water, ethylene glycol, the ionic liquid [EMIM][TCM], and copper, with relative standard deviations averaging less than 1%. Measurements on [EMIM][DCA] indicated a systematic deviation from the literature values. The study highlights the repeatability of the measurement method, which worked well for both temperature increases and decreases, as well as for the mean of the two. The influence of sample size on the results when it came to liquids was also investigated, revealing that large sample sizes led to underpredictions; while, small sizes yielded the opposite effect. The best results were obtained with half-filled vials; a similar filling level as was used in the electrical calibrations of the calorimeters with heaters in the vials. Additionally, no significant differences were observed among the eight calorimeters of the I-Cal Flex instrument, and different baseline calculation methods had negligible impact on the results. Overall, this study illustrates how a heat conduction calorimeter can be utilized for accurate and precise heat capacity measurements for both solid and liquid substances.

(Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Calibration, Ionic liquids, Isothermal calorimetry
in
Journal of Thermal Analysis and Calorimetry
volume
149
issue
5
pages
10 pages
publisher
Akademiai Kiado
external identifiers
  • scopus:85185261568
ISSN
1388-6150
DOI
10.1007/s10973-023-12824-8
language
English
LU publication?
yes
id
cc8b288c-cf72-444a-b48c-82c50dd0d60b
date added to LUP
2024-03-26 14:05:03
date last changed
2024-03-26 14:05:35
@article{cc8b288c-cf72-444a-b48c-82c50dd0d60b,
  abstract     = {{<p>This study investigates the suitability of heat conduction calorimeters for determining the specific heat capacity of solid and liquid substances. Accurate and precise measurements were obtained for various substances, including water, ethylene glycol, the ionic liquid [EMIM][TCM], and copper, with relative standard deviations averaging less than 1%. Measurements on [EMIM][DCA] indicated a systematic deviation from the literature values. The study highlights the repeatability of the measurement method, which worked well for both temperature increases and decreases, as well as for the mean of the two. The influence of sample size on the results when it came to liquids was also investigated, revealing that large sample sizes led to underpredictions; while, small sizes yielded the opposite effect. The best results were obtained with half-filled vials; a similar filling level as was used in the electrical calibrations of the calorimeters with heaters in the vials. Additionally, no significant differences were observed among the eight calorimeters of the I-Cal Flex instrument, and different baseline calculation methods had negligible impact on the results. Overall, this study illustrates how a heat conduction calorimeter can be utilized for accurate and precise heat capacity measurements for both solid and liquid substances.</p>}},
  author       = {{Hothar, Marcus and Wadsö, Lars}},
  issn         = {{1388-6150}},
  keywords     = {{Calibration; Ionic liquids; Isothermal calorimetry}},
  language     = {{eng}},
  number       = {{5}},
  pages        = {{2179--2188}},
  publisher    = {{Akademiai Kiado}},
  series       = {{Journal of Thermal Analysis and Calorimetry}},
  title        = {{Accurate heat capacity determination of solids and liquids using a heat conduction calorimeter}},
  url          = {{http://dx.doi.org/10.1007/s10973-023-12824-8}},
  doi          = {{10.1007/s10973-023-12824-8}},
  volume       = {{149}},
  year         = {{2024}},
}