Automatic Intrinsic Calibration of Double-sided Silicon Strip Detectors

Reese, M. ; Gerl, J. ; Golubev, Pavel LU and Pietralla, N. (2015) In Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment 779. p.63-68
Abstract
Abstract A reliable and simple-to-use algorithm was developed for the energy-calibration of double-sided silicon strip detectors (DSSSDs). It works by enforcing mutual consistency of p-side and n-side information for every detected event. The procedure does not rely on a dedicated data set for calibration and is robust enough to work fully automated without human supervision. The method was developed and applied to data from a DSSSD of the Lund-York-Cologne CAlorimeter (LYCCA) for the HISPEC experiment at FAIR. It has been tested on ions in the A â 90 mass range at energies of E kin â 300 MeV / u .
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Silicon strip detector, Calibration procedure, Energy loss measurement
in
Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment
volume
779
pages
63 - 68
publisher
Elsevier
external identifiers
  • wos:000349672600010
  • scopus:84922249871
ISSN
0167-5087
DOI
10.1016/j.nima.2015.01.032
language
English
LU publication?
yes
id
a5c0ce9a-e33f-4af2-88c1-03b521361e53 (old id 5043370)
alternative location
http://www.sciencedirect.com/science/article/pii/S0168900215000637
date added to LUP
2016-04-01 14:33:38
date last changed
2022-04-14 18:28:23
@article{a5c0ce9a-e33f-4af2-88c1-03b521361e53,
  abstract     = {{Abstract A reliable and simple-to-use algorithm was developed for the energy-calibration of double-sided silicon strip detectors (DSSSDs). It works by enforcing mutual consistency of p-side and n-side information for every detected event. The procedure does not rely on a dedicated data set for calibration and is robust enough to work fully automated without human supervision. The method was developed and applied to data from a DSSSD of the Lund-York-Cologne CAlorimeter (LYCCA) for the HISPEC experiment at FAIR. It has been tested on ions in the A â 90 mass range at energies of E kin â 300 MeV / u .}},
  author       = {{Reese, M. and Gerl, J. and Golubev, Pavel and Pietralla, N.}},
  issn         = {{0167-5087}},
  keywords     = {{Silicon strip detector; Calibration procedure; Energy loss measurement}},
  language     = {{eng}},
  pages        = {{63--68}},
  publisher    = {{Elsevier}},
  series       = {{Nuclear Instruments & Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors, and Associated Equipment}},
  title        = {{Automatic Intrinsic Calibration of Double-sided Silicon Strip Detectors}},
  url          = {{https://lup.lub.lu.se/search/files/4037610/5043378.pdf}},
  doi          = {{10.1016/j.nima.2015.01.032}},
  volume       = {{779}},
  year         = {{2015}},
}