Aneurysm clip MR artifacts. Titanium versus stainless steel and influence of imaging parameters

Björkman-Burtscher, Isabella LU ; Owman, Titti LU ; Romner, Bertil LU ; Ståhlberg, Freddy LU and Holtås, Stig LU (1998) In Acta Radiologica 39(1). p.70-76
Abstract
PURPOSE: The aim of this study was to evaluate the extent to which titanium aneurysm clips could improve the quality of MR imaging compared with stainless steel clips, and to determine whether the clip artifacts could be reduced by controlling certain MR imaging parameters in frequently used pulse sequences. MATERIAL AND METHODS: The metal artifacts induced by 3 aneurysm clips were compared in 3 pulse sequences. The clips were: a Yasargil titanium aneurysm clip FT 752 T; a Yasargil standard aneurysm clip FE 752 K; and, for comparison, a ferromagnetic Scoville aneurysm clip En-58J. The pulse sequences were: spin echo (SE); gradient echo (GE); and fast SE. An evaluation was made of 3 imaging parameters with regard to their influence on the... (More)
PURPOSE: The aim of this study was to evaluate the extent to which titanium aneurysm clips could improve the quality of MR imaging compared with stainless steel clips, and to determine whether the clip artifacts could be reduced by controlling certain MR imaging parameters in frequently used pulse sequences. MATERIAL AND METHODS: The metal artifacts induced by 3 aneurysm clips were compared in 3 pulse sequences. The clips were: a Yasargil titanium aneurysm clip FT 752 T; a Yasargil standard aneurysm clip FE 752 K; and, for comparison, a ferromagnetic Scoville aneurysm clip En-58J. The pulse sequences were: spin echo (SE); gradient echo (GE); and fast SE. An evaluation was made of 3 imaging parameters with regard to their influence on the size of the metal artifacts. The parameters were: bandwidth; echo time (TE); and echo-train length. RESULTS: The titanium clip showed artifacts that were about 60% smaller than those from the stainless steel clip. The only parameter that influenced artifact size to any major degree was bandwidth in the SE sequences but not in the GE sequences. GE sequences induced larger artifacts than SE sequences and showed larger artifacts with longer TE. CONCLUSION: Titanium aneurysm clips reduced MR artifacts by approximately 60% compared to stainless steel clips. Artifacts were further reduced by using SE-based sequences with a high bandwidth or, if necessary, GE sequences with a low TE. (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Acta Radiologica
volume
39
issue
1
pages
70 - 76
publisher
SAGE Publications
external identifiers
  • pmid:9498874
  • scopus:0031602920
ISSN
1600-0455
language
English
LU publication?
yes
id
70f71e6a-aab7-444f-9f4c-70ab1d0fdcb6 (old id 1113031)
date added to LUP
2016-04-01 17:03:00
date last changed
2022-01-29 00:01:01
@article{70f71e6a-aab7-444f-9f4c-70ab1d0fdcb6,
  abstract     = {{PURPOSE: The aim of this study was to evaluate the extent to which titanium aneurysm clips could improve the quality of MR imaging compared with stainless steel clips, and to determine whether the clip artifacts could be reduced by controlling certain MR imaging parameters in frequently used pulse sequences. MATERIAL AND METHODS: The metal artifacts induced by 3 aneurysm clips were compared in 3 pulse sequences. The clips were: a Yasargil titanium aneurysm clip FT 752 T; a Yasargil standard aneurysm clip FE 752 K; and, for comparison, a ferromagnetic Scoville aneurysm clip En-58J. The pulse sequences were: spin echo (SE); gradient echo (GE); and fast SE. An evaluation was made of 3 imaging parameters with regard to their influence on the size of the metal artifacts. The parameters were: bandwidth; echo time (TE); and echo-train length. RESULTS: The titanium clip showed artifacts that were about 60% smaller than those from the stainless steel clip. The only parameter that influenced artifact size to any major degree was bandwidth in the SE sequences but not in the GE sequences. GE sequences induced larger artifacts than SE sequences and showed larger artifacts with longer TE. CONCLUSION: Titanium aneurysm clips reduced MR artifacts by approximately 60% compared to stainless steel clips. Artifacts were further reduced by using SE-based sequences with a high bandwidth or, if necessary, GE sequences with a low TE.}},
  author       = {{Björkman-Burtscher, Isabella and Owman, Titti and Romner, Bertil and Ståhlberg, Freddy and Holtås, Stig}},
  issn         = {{1600-0455}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{70--76}},
  publisher    = {{SAGE Publications}},
  series       = {{Acta Radiologica}},
  title        = {{Aneurysm clip MR artifacts. Titanium versus stainless steel and influence of imaging parameters}},
  volume       = {{39}},
  year         = {{1998}},
}