A miniaturised electrochemical affinity assay based on a wall-free sample droplet and micro-dispensing of the redox-labelled binding partner
(2001) In Biosensors & Bioelectronics 16(9-12). p.611-620- Abstract
- An affinity-assay was developed that is based on the modulation of the diffusion coefficient of a redox-labelled hapten upon complementary recognition of the analyte leading to an increase of molecular weight and hence to a decrease of the diffusion coefficient. The slower diffusion is monitored by means of cyclic voltammetry. In order to demonstrate the feasibility of this assay format, recognition of biotin by streptavidin has been chosen as a model system. Labelling of biotin was achieved by covalent binding of a ferrocene derivative to the biotin unit. To reduce the consumption of expensive compounds and to allow automatisation of the assay a novel miniaturised set-up was developed based on a wall-free sample droplet which forms the... (More)
- An affinity-assay was developed that is based on the modulation of the diffusion coefficient of a redox-labelled hapten upon complementary recognition of the analyte leading to an increase of molecular weight and hence to a decrease of the diffusion coefficient. The slower diffusion is monitored by means of cyclic voltammetry. In order to demonstrate the feasibility of this assay format, recognition of biotin by streptavidin has been chosen as a model system. Labelling of biotin was achieved by covalent binding of a ferrocene derivative to the biotin unit. To reduce the consumption of expensive compounds and to allow automatisation of the assay a novel miniaturised set-up was developed based on a wall-free sample droplet which forms the electrochemical cell with typical volumes of up to 10 mul. This droplet is dispensed by means of a step-motor driven syringe pump through a specially designed electrode holder spanning the gap between a micro-working electrode and a macroscopic counter electrode. By means of a piezo-driven micro-dispenser a predefined number of nano-droplets (100 pl volume each) containing the redox-labelled hapten are shot into the sample droplet. By this, any physical contact and hence any cross-contamination between the sample and the reagent solution could be avoided. Signal amplification can be achieved by redox recycling between the micro-electrode and the perpendicular positioned macroscopic counter electrode. (C) 2001 Elsevier Science B.V. All rights reserved. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/2376417
- author
- Mosbach, M. ; Laurell, Thomas LU ; Nilsson, J. ; Csöregi, Elisabeth LU and Schuhmann, W.
- organization
- publishing date
- 2001
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- AVIDIN, TECHNOLOGY, MALDI, BIOTIN INTERACTION, CAPILLARY-ELECTROPHORESIS, FLOW-THROUGH DISPENSER, streptavidin, miniaturisation, micro-dispenser, immunoassay, biotin, diffusion coefficient
- in
- Biosensors & Bioelectronics
- volume
- 16
- issue
- 9-12
- pages
- 611 - 620
- publisher
- Elsevier
- external identifiers
-
- wos:000172169500002
- scopus:0034777036
- ISSN
- 1873-4235
- DOI
- 10.1016/s0956-5663(01)00186-5
- language
- English
- LU publication?
- yes
- additional info
- The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Analytical Chemistry (S/LTH) (011001004), Biomedical Engineering (011200011)
- id
- 7539ee43-ad99-4ce7-97ef-de2b67eaf944 (old id 2376417)
- date added to LUP
- 2016-04-01 16:52:08
- date last changed
- 2022-03-30 18:54:00
@article{7539ee43-ad99-4ce7-97ef-de2b67eaf944, abstract = {{An affinity-assay was developed that is based on the modulation of the diffusion coefficient of a redox-labelled hapten upon complementary recognition of the analyte leading to an increase of molecular weight and hence to a decrease of the diffusion coefficient. The slower diffusion is monitored by means of cyclic voltammetry. In order to demonstrate the feasibility of this assay format, recognition of biotin by streptavidin has been chosen as a model system. Labelling of biotin was achieved by covalent binding of a ferrocene derivative to the biotin unit. To reduce the consumption of expensive compounds and to allow automatisation of the assay a novel miniaturised set-up was developed based on a wall-free sample droplet which forms the electrochemical cell with typical volumes of up to 10 mul. This droplet is dispensed by means of a step-motor driven syringe pump through a specially designed electrode holder spanning the gap between a micro-working electrode and a macroscopic counter electrode. By means of a piezo-driven micro-dispenser a predefined number of nano-droplets (100 pl volume each) containing the redox-labelled hapten are shot into the sample droplet. By this, any physical contact and hence any cross-contamination between the sample and the reagent solution could be avoided. Signal amplification can be achieved by redox recycling between the micro-electrode and the perpendicular positioned macroscopic counter electrode. (C) 2001 Elsevier Science B.V. All rights reserved.}}, author = {{Mosbach, M. and Laurell, Thomas and Nilsson, J. and Csöregi, Elisabeth and Schuhmann, W.}}, issn = {{1873-4235}}, keywords = {{AVIDIN; TECHNOLOGY; MALDI; BIOTIN INTERACTION; CAPILLARY-ELECTROPHORESIS; FLOW-THROUGH DISPENSER; streptavidin; miniaturisation; micro-dispenser; immunoassay; biotin; diffusion coefficient}}, language = {{eng}}, number = {{9-12}}, pages = {{611--620}}, publisher = {{Elsevier}}, series = {{Biosensors & Bioelectronics}}, title = {{A miniaturised electrochemical affinity assay based on a wall-free sample droplet and micro-dispensing of the redox-labelled binding partner}}, url = {{http://dx.doi.org/10.1016/s0956-5663(01)00186-5}}, doi = {{10.1016/s0956-5663(01)00186-5}}, volume = {{16}}, year = {{2001}}, }