Comparison between software volumetric breast density estimates in breast tomosynthesis and digital mammography images in a large public screening cohort

Förnvik, Daniel LU ; Förnvik, Hannie LU ; Fieselmann, Andreas ; Lång, Kristina LU and Sartor, Hanna LU (2019) In European Radiology 29(1). p.330-336
Abstract

Objectives: To compare software estimates of volumetric breast density (VBD) based on breast tomosynthesis (BT) projections to those based on digital mammography (DM) images in a large screening cohort, the Malmö Breast Tomosynthesis Screening Trial (MBTST). Methods: DM and BT images of 9909 women (enrolled 2010–2015) were retrospectively analysed with prototype software to estimate VBD. Software calculation is based on a physics model of the image acquisition process and incorporates the effect of masking in DM based on accumulated dense tissue areas. VBD (continuously and categorically) was compared between BT [central projection (mediolateral oblique view (MLO)] and two-view DM, and with radiologists’ BI-RADS density 4th... (More)

Objectives: To compare software estimates of volumetric breast density (VBD) based on breast tomosynthesis (BT) projections to those based on digital mammography (DM) images in a large screening cohort, the Malmö Breast Tomosynthesis Screening Trial (MBTST). Methods: DM and BT images of 9909 women (enrolled 2010–2015) were retrospectively analysed with prototype software to estimate VBD. Software calculation is based on a physics model of the image acquisition process and incorporates the effect of masking in DM based on accumulated dense tissue areas. VBD (continuously and categorically) was compared between BT [central projection (mediolateral oblique view (MLO)] and two-view DM, and with radiologists’ BI-RADS density 4th ed. scores. Agreement and correlation were investigated with weighted kappa (κ), Spearman’s correlation coefficient (r), and Bland–Altman analysis. Results: There was a high correlation (r = 0.83) between VBD in DM and BT and substantial agreement between the software breast density categories [observed agreement, 61.3% and 84.8%; κ = 0.61 and ĸ = 0.69 for four (a/b/c/d) and two (fat involuted vs. dense) density categories, respectively]. There was moderate agreement between radiologists’ BI-RADS scores and software density categories in DM (ĸ = 0.55) and BT (ĸ = 0.47). Conclusions: In a large public screening setting, we report a substantial agreement between VBD in DM and BT using software with special focus on masking effect. This automated and objective mode of measuring VBD may be of value to radiologists and women when BT is used as the primary breast cancer screening modality. Key Points: • There was a high correlation between continuous volumetric breast density in DM and BT.• There was substantial agreement between software breast density categories (four groups) in DM and BT; with clinically warranted binary software breast density categories, the agreement increased markedly.• There was moderate agreement between radiologists’ BI-RADS scores and software breast density categories in DM and BT.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Breast neoplasms, Diagnostic imaging, Digital breast tomosynthesis, Mammography, Mass screening
in
European Radiology
volume
29
issue
1
pages
330 - 336
publisher
Springer
external identifiers
  • pmid:29943180
  • scopus:85049007800
ISSN
0938-7994
DOI
10.1007/s00330-018-5582-0
language
English
LU publication?
yes
id
82f10515-e9c8-4744-832e-786eef84460b
date added to LUP
2018-07-04 13:39:27
date last changed
2024-02-13 22:46:40
@article{82f10515-e9c8-4744-832e-786eef84460b,
  abstract     = {{<p>Objectives: To compare software estimates of volumetric breast density (VBD) based on breast tomosynthesis (BT) projections to those based on digital mammography (DM) images in a large screening cohort, the Malmö Breast Tomosynthesis Screening Trial (MBTST). Methods: DM and BT images of 9909 women (enrolled 2010–2015) were retrospectively analysed with prototype software to estimate VBD. Software calculation is based on a physics model of the image acquisition process and incorporates the effect of masking in DM based on accumulated dense tissue areas. VBD (continuously and categorically) was compared between BT [central projection (mediolateral oblique view (MLO)] and two-view DM, and with radiologists’ BI-RADS density 4<sup>th</sup> ed. scores. Agreement and correlation were investigated with weighted kappa (κ), Spearman’s correlation coefficient (r), and Bland–Altman analysis. Results: There was a high correlation (r = 0.83) between VBD in DM and BT and substantial agreement between the software breast density categories [observed agreement, 61.3% and 84.8%; κ = 0.61 and ĸ = 0.69 for four (a/b/c/d) and two (fat involuted vs. dense) density categories, respectively]. There was moderate agreement between radiologists’ BI-RADS scores and software density categories in DM (ĸ = 0.55) and BT (ĸ = 0.47). Conclusions: In a large public screening setting, we report a substantial agreement between VBD in DM and BT using software with special focus on masking effect. This automated and objective mode of measuring VBD may be of value to radiologists and women when BT is used as the primary breast cancer screening modality. Key Points: • There was a high correlation between continuous volumetric breast density in DM and BT.• There was substantial agreement between software breast density categories (four groups) in DM and BT; with clinically warranted binary software breast density categories, the agreement increased markedly.• There was moderate agreement between radiologists’ BI-RADS scores and software breast density categories in DM and BT.</p>}},
  author       = {{Förnvik, Daniel and Förnvik, Hannie and Fieselmann, Andreas and Lång, Kristina and Sartor, Hanna}},
  issn         = {{0938-7994}},
  keywords     = {{Breast neoplasms; Diagnostic imaging; Digital breast tomosynthesis; Mammography; Mass screening}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{330--336}},
  publisher    = {{Springer}},
  series       = {{European Radiology}},
  title        = {{Comparison between software volumetric breast density estimates in breast tomosynthesis and digital mammography images in a large public screening cohort}},
  url          = {{http://dx.doi.org/10.1007/s00330-018-5582-0}},
  doi          = {{10.1007/s00330-018-5582-0}},
  volume       = {{29}},
  year         = {{2019}},
}